DigitalFlow™ XMT868i Service Manual
21
Appendix B. Calibration and Testing with PanaView
1.
From the
Calibration/Test
menu (see Figure 11 on page 20), double-click on the
Calibration
entry to open the
option.
2. In the center pane, double-click on Slot 0.
Note:
The
Slot 1
option only appears in the center pane if an option card has been installed in this slot.Double-click
on either
Output A
or
Output B
to open the Analog Output menu. (
Output A
will be used as an example in this
manual.)
IMPORTANT:
The procedure for calibrating
Output B
is identical to that for calibrating
Output A
. However, when
calibrating
Output B
, be sure to reconnect the ammeter on terminal block
J1
. See Figure 12 on page 20
for the correct pin numbers.
3. Double-click on
4 milliamps
to calibrate the low end of the output range.
4. Double-click on
4 mA UP
or
4 mA DOWN
to adjust the ammeter reading UP or DOWN, until a 4 mA reading is
achieved, or
Double-click on
4 mA Numer,
and directly enter the mA reading in the right pane.
5. Once you have achieved a 4 mA reading, double-click on either
4 mA STORE
to store the calibration, or on
4 mA
ABORT
to exit the menu without storing the calibration.
Note:
If the ammeter reading cannot be adjusted within 5.0
A of the 4 or 20 mA setting, contact the factory for
assistance.
6. PanaView returns to the
Analog Output
window. Double-click on
20 milliamps
, and repeat steps 5 and 6 to cali
-
brate the high end of the output range.
Note:
Skip the following prompt, if linearity testing of the output will not be performed at this time.
7. Double-click on
% Test
to test the linearity of the currently selected analog output.
8. Check the ammeter reading at
50% output
. Then, enter a different
output
(0-100%) and double-click on
percent
.
Check the ammeter reading at this setting and click [Exit Page] when done.
Table 8 on page 21 lists the expected ammeter readings at various
% Full Scale
settings, for both 4-20 mA and
0-20 mA scales. Refer to this table to verify the accuracy of the ammeter readings taken above.
If the linearity test readings are not within 5
A of the values listed in Table 8, check the accuracy and wiring of the
ammeter. Then, repeat the low and high end calibrations. If the analog output still does not pass the linearity test,
contact the factory for assistance.
9. Double-click on
Exit Function
to return to the
Calibration
menu. Then choose another output to calibrate or click
on [Exit Page] to leave the
Calibration/Test
menu.
This completes calibration of the Slot 0 analog outputs. Proceed to the appropriate section to calibrate additional
inputs/outputs.
Table 8: Expected Ammeter Readings
% Full Scale
4-20 mA Scale*
0-20 mA Scale*
0
4.000
0.000
10
5.600
2.000
20
7.200
4.000
30
8.800
6.000
40
10.400
8.000
50
12.000
10.000
60
13.600
12.000
70
15.200
14.000
80
16.800
16.000
90
18.400
18.000
100
20.000
20.000
* all ammeter readings should be ±0.005 mA
Содержание DigitalFlow XMT868i
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 24: ...Chapter 2 Error Codes 4 DigitalFlow XMT868i Service Manual...
Страница 34: ...Chapter 3 Diagnostics 10 DigitalFlow XMT868i Service Manual...
Страница 44: ...Chapter 4 Parts Replacement 10 DigitalFlow XMT868i Service Manual...
Страница 48: ...Chapter 4 Parts Replacement DigitalFlow XMT868i Service Manual...
Страница 52: ...Appendix A Service Record 18 DigitalFlow XMT868i Service Manual...
Страница 66: ...Appendix B Calibration and Testing with PanaView 32 DigitalFlow XMT868i Service Manual...
Страница 72: ...Warranty 38 DigitalFlow XMT868i Service Manual no content intended for this page...
Страница 73: ......