Chapter 1. Calibration
4
DigitalFlow™ XMT868i Service Manual
1.3.4
Testing Output Linearity (cont.)
Table 1 lists the expected ammeter readings at various
% Full Scale
settings, for both 4-20 mA and 0-20 mA scales.
Refer to this table to verify the accuracy of the ammeter readings taken above.
If the linearity test readings are not within 16
A of the values listed in Table 1, check the accuracy and wiring of the
ammeter. Then, repeat the low and high end calibrations. If the analog output still does not pass the linearity test,
contact the factory for assistance.
10.
Scroll to
EXIT
and press
[Enter]
to return to the
Cal
menu. Then choose another output to calibrate or press
[Escape]
to leave the
CALIB
menu.
This completes calibration of the
Slot 0
analog outputs. Proceed to the appropriate section to calibrate additional
inputs/outputs.
1.4 Calibrating Analog Inputs
Analog inputs may be added to the Model XMT868i by installing an appropriate option card in
Slot 1
. This option card
contains two or four analog inputs, which are designated as
A, B, C
and
D
. Both the zero-point and full-scale values
for each input must be calibrated.
Calibration of the analog inputs requires the use of a calibrated current source. If an independent calibrated current
source is not available, one of the
Slot 0
analog outputs may be used for the calibration. During the analog input
calibration, the
Slot 0
analog output will supply the low reference, high reference, 4 mA and 20 mA signals at the
appropriate times.
IMPORTANT:
If a
Slot 0
analog output will be used to calibrate the analog inputs, the
Slot 0
analog output calibration
procedure must be completed first.
Table 1: Expected Ammeter Readings
% Full Scale
4-20 mA Scale*
0-20 mA Scale*
0
4.000
0.000
10
5.600
2.000
20
7.200
4.000
30
8.800
6.000
40
10.400
8.000
50
12.000
10.000
60
13.600
12.000
70
15.200
14.000
80
16.800
16.000
90
18.400
18.000
100
20.000
20.000
* all ammeter readings should be ±0.016 mA
Содержание DigitalFlow XMT868i
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 24: ...Chapter 2 Error Codes 4 DigitalFlow XMT868i Service Manual...
Страница 34: ...Chapter 3 Diagnostics 10 DigitalFlow XMT868i Service Manual...
Страница 44: ...Chapter 4 Parts Replacement 10 DigitalFlow XMT868i Service Manual...
Страница 48: ...Chapter 4 Parts Replacement DigitalFlow XMT868i Service Manual...
Страница 52: ...Appendix A Service Record 18 DigitalFlow XMT868i Service Manual...
Страница 66: ...Appendix B Calibration and Testing with PanaView 32 DigitalFlow XMT868i Service Manual...
Страница 72: ...Warranty 38 DigitalFlow XMT868i Service Manual no content intended for this page...
Страница 73: ......