Chapter 1. Calibration
2
DigitalFlow™ XMT868i Service Manual
1.3
Calibrating and Testing Analog Outputs
Every Model XMT868i flowmeter includes two built-in analog outputs (A and B) at terminal block
J1
, which is
designated as
Slot 0
. Additional analog outputs may be added to the Model XMT868i by installing an appropriate
option card in
Slot 1
. This option card contains two analog outputs, which are also designated as
A
and
B
.
Before beginning calibration, an ammeter must be connected to the desired analog output. Both the zero-point and
full-scale values for all of these outputs must be calibrated. After calibrating the outputs, which have a resolution of
16.0
A (0.1% full scale), their linearity should be tested.
Note:
The zero point of the analog output may be set at either 0 mA or 4 mA. However, the calibration always uses
the 4 mA point, as the meter extrapolates this value to obtain the 0 mA point.
1.3.1
Preparing for Calibration
Prepare for calibration by inserting an ammeter in series with the load on the desired output. Refer to Figure 1 and
Figure 4 on page 12 to identify the
OUT(+)
and
RTN(-)
pins for the desired output.
Figure 1: Ammeter Connection (Output A)
Terminal Block J1 (Slot 0)
B
RTN
B
SIG
A
RTN
A
SIG
4
1
+
-
Ammeter
Load
Содержание DigitalFlow XMT868i
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 24: ...Chapter 2 Error Codes 4 DigitalFlow XMT868i Service Manual...
Страница 34: ...Chapter 3 Diagnostics 10 DigitalFlow XMT868i Service Manual...
Страница 44: ...Chapter 4 Parts Replacement 10 DigitalFlow XMT868i Service Manual...
Страница 48: ...Chapter 4 Parts Replacement DigitalFlow XMT868i Service Manual...
Страница 52: ...Appendix A Service Record 18 DigitalFlow XMT868i Service Manual...
Страница 66: ...Appendix B Calibration and Testing with PanaView 32 DigitalFlow XMT868i Service Manual...
Страница 72: ...Warranty 38 DigitalFlow XMT868i Service Manual no content intended for this page...
Страница 73: ......