DigitalFlow™ XGF868i Service Manual
17
Chapter 3. Diagnostics
Chapter 3. Diagnostics
3.1
Introduction
This chapter explains how to troubleshoot the XGF868i if problems arise with the electronics enclosure, the flowcell, or
the transducers. Indications of a possible problem include:
• Display of an error message on the LCD display screen
• Erratic flow readings
• Readings of doubtful accuracy (i.e., readings that are not consistent with readings from another flow
measuring device connected to the same process).
If any of the above conditions occurs, proceed with the instructions presented in this chapter.
3.2 Displaying Diagnostic Parameters
The XGF868i has built-in
Diagnostic Parameters
to aid in the troubleshooting of flowcell, transducer and/or electrical
problems. To access these parameters, use the
Keypad Program
. Then, complete the following instructions to display
the desired diagnostic parameter:
Note:
To perform this function with PanaView software see Appendix C,
Programming with PanaView
in the
Programming Manual,
and/or the PanaView
User’s Manual
.
1.
Press
[Escape], [Enter], [Escape].
2.
In the
Keypad Program
window, scroll to
PROG
and press
[
Enter].
3.
In the
PROG
menu, scroll to
GLOBL
and press
[Enter].
4.
Scroll to
I/O
and press
[Enter].
5.
Scroll to
LCD
and press
[Enter].
6.
The window now asks for the
# of LCD Parameters
. Scroll to the desired number (from
OFF
through 1-4 and
KEY
)
and press
[Enter]
.
The
OFF
setting switches the measurement display off, while the
KEY
setting enables users to change the
measurement display via the arrow keys, without accessing the
Keypad Program.
If you select
KEY:
•
To view a parameter other than the one currently displayed, press the [r] or [s] keys to scroll through the
various parameters.
•
To scroll through the channel options on an XGF868i, press the [v] or [w] keys until you have reached the
desired option.
7.
Scroll to the desired
Channel option
, as listed in
Table 2
below.
8.
For each channel, select the desired
Measurement Parameter
, as shown in
Note:
The measurement units that appear in these prompts are those selected in the
GLOBL-SYSTM
menu earlier
in this section. Also, when differences in one channel's programming invalidate an output previously chosen
for the other, the measurement defaults to the nearest selectable item in the parameter list.
The previous two prompts repeat until all of the specified
# of LCD Parameters
have been set up. When all of the
display parameters have been set up, the meter returns to the
Global I/O
window. To leave the
Keypad Program
,
press the
[Escape]
key three times.
Table 2: Channel Options
Option
Description
CH1
Channel 1
CH2
Channel 2
SUM
CH1+CH2
DIF
CH1-CH2
AVE
(CH1+CH2)/2
Содержание DigitalFlow XGF868i
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 24: ...Chapter 2 Error Codes 16 DigitalFlow XGF868i Service Manual no content intended for this page...
Страница 40: ...Chapter 4 Parts Replacement DigitalFlow XGF868i Service Manual 32...
Страница 44: ...Appendix A Service Record 36 DigitalFlow XGF868i Service Manual no content intended for this page...
Страница 70: ...Warranty 62 DigitalFlow XGF868i Service Manual no content intended for this page...
Страница 71: ......