Chapter 1. Calibration
2
DigitalFlow™ XGF868i Service Manual
1.3
Calibrating and Testing the Analog Outputs
Every XGF868i flowmeter includes two built-in analog outputs (A and B) at terminal block
J1
, which is designated as
Slot 0
. Additional analog outputs may be added to the Model XGF868i by installing an appropriate option card in
Slot 1
. This option card contains two analog outputs, which are also designated as A and B.
Before beginning calibration, an ammeter must be connected to the desired analog output. Both the zero-point and
full-scale values for all of these outputs must be calibrated. After calibrating the outputs, which have a resolution of
5.0
A (0.03% full scale), their linearity should be tested.
Note:
The zero point of the analog output may be set at either 0 mA or 4 mA. However, the calibration always uses
the 4 mA point, as the meter extrapolates this value to obtain the 0 mA point.
1.3.1
Preparing for Calibration
Prepare for calibration by inserting an ammeter in series with the load on the desired output. Refer to
below
to identify the
OUT(+)
and
RTN(-)
pins for the desired output.
Figure 1: Ammeter Connection (Output A)
Terminal Block J1 (Slot 0)
B
RTN
B
SIG
A
RTN
A
SIG
4
1
+
-
Ammeter
Load
Содержание DigitalFlow XGF868i
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 24: ...Chapter 2 Error Codes 16 DigitalFlow XGF868i Service Manual no content intended for this page...
Страница 40: ...Chapter 4 Parts Replacement DigitalFlow XGF868i Service Manual 32...
Страница 44: ...Appendix A Service Record 36 DigitalFlow XGF868i Service Manual no content intended for this page...
Страница 70: ...Warranty 62 DigitalFlow XGF868i Service Manual no content intended for this page...
Страница 71: ......