Chapter 5. Clearing Data
86
DigitalFlow™ DF868 Multipurpose Ultrasonic Liquid Flowmeter Programming Manual (2-Channel)
5.3 Deleting Site Files
Use the
SITE
submenu to clear site files from the DF868’s memory. Use the
[SCREEN]
key to activate the desired
display pane, and refer to the menu map in Appendix A, page A-8 while completing the following steps:
1.
To access the
Clear Menu
, press the
[CLR]
key.
a.
If you have activated the security feature (see Chapter 1,
Programming Site Data,
page 50 for details), enter the
assigned password and press
[ENT]
.
2.
Press
[F2]
=
SITE
.
3.
At the
Site Name
prompt, use the [
, [
and
[F1]-[F4]
keys to select a site file for deletion. To leave the
SITE
submenu, press the
[EXIT]
key.
Note:
At the
SITE NAME
prompt, the option bar shows the names of all site files currently stored in memory.
4.
Press
[F1] to abort the procedure
, or press
[F2] to Clear the specified site file. If any site files remain in memory,
the Site Name prompt repeats.
When all of the stored site files have been cleared or the
[EXIT]
key has been pressed at the
SITE NAME
prompt, the
programming sequence resumes at this point.
5.4 Deleting Log Files
Use the
LOG
submenu to clear log files from the Model DF868’s memory. Press the appropriate side of the
[SCREEN]
key to activate the desired pane of the data display screen, and complete the following steps:
1.
To access the
Clear Menu
, press the
[CLR]
key.
a.
If you have activated the security feature (see Chapter 1,
Programming Site Data,
page 50 for details), enter the
assigned password and press
[ENT]
.
2.
Press
[F3]
=
LOG
.
3.
At the
Name
prompt, the option bar shows the names of all log files currently stored in memory. Use the [
, [
and
[F1]-[F4]
keys to select a site file for deletion. To leave the
LOG
submenu, press the
[EXIT]
key.
4.
Press
[F1]
to abort the procedure or press
[F2]
to
Clear
the specified log file. If any log files remain in memory, the
Name
prompt repeats.
When all of the stored log files have been cleared or the
[EXIT]
key has been pressed at the
NAME
prompt, the
programming sequence resumes at this point.
Procedure Options
Do one of the following:
•
Press
[F1]-[F3]
to enter one of the submenus.
•
Press the
[EXIT]
key to return to the data measurement display.
Содержание DigitalFlow DF868
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 10: ...Contents viii DigitalFlow DF868 Programming Manual 2 Channel...
Страница 104: ...Appendix A Menu Maps 96 DigitalFlow DF868 Multipurpose Ultrasonic Liquid Flowmeter Programming Manual 2 Channel...
Страница 118: ...Appendix B Data Records 110 DigitalFlow DF868 Multipurpose Ultrasonic Liquid Flowmeter Programming Manual 2 Channel...
Страница 150: ...Appendix E OPC Server 142 DigitalFlow DF868 Multipurpose Ultrasonic Liquid Flowmeter Programming Manual 2 Channel...
Страница 158: ...2 DigitalFlow DF868 Multipurpose Ultrasonic Liquid Flowmeter Programming Manual 2 Channel...
Страница 159: ......