Chapter 3. Logging Data
72
DigitalFlow™ DF868 Multipurpose Ultrasonic Liquid Flowmeter Programming Manual (2-Channel)
3.4 Creating an ERROR Log
Use the
ERROR
submenu to create a new error log and to select the logging parameters. An error log updates every 5
seconds (when the display updates), but only if a new error condition occurs. Error logs have a fixed length of 2 pages
and contain sixty records per page. Each record shows the time of the error, the measurement parameter values at
that time, and the error code message. The logged values of the chosen measurement parameters at the time of the
error condition provide valuable troubleshooting information.
Press the appropriate side of the
[SCREEN]
key to activate the desired pane of the display screen, and complete the
following steps:
1.
To access the
Log Menu
, press
[LOG]
.
a.
If you have activated the security feature (see Chapter 1,
Programming Site Data
, page 50 for details), enter the
assigned password and press
[ENT]
.
2.
Press
[F4]
=
ERROR.
3.
Use the [
, [
and
[F1]-[F4]
keys to inspect a completed log or to change an active log, or enter a new
Name
(up to 5 characters) and press
[ENT]
to create a new log file.
Note:
The option bar lists all error logs currently stored in memory. Recall that a flashing asterisk (*) appears at
the far right of the locator bar if the Model DF868 is currently logging data (see Chapter 3,
Operation
, page
3-4 of the
Startup Guide
).
4.
At the
Log Message
prompt, enter a brief description (up to 21 characters) of the log and press
[ENT]
.
5.
Use the [
, [
and
[F1]-[F4]
keys to select the first parameter to be logged. See Table 36 for a list of the
available options.
IMPORTANT:
Only the first parameter logged may be graphically displayed (see page 63).
Table 36: Measurement Parameter Options
Option Bar Choice
Description
[F1]
=
VEL
Flow Velocity
[F2]
=
VOLUM
Volumetric Flow
[F3]
=
+TOTL
Forward Totalized Volume Flow
[F4]
=
-TOTL
Reverse Totalized Volume Flow
[
+ [F1]
=
MDOT*
Mass Flow
[
+ [F2]
=
+MASS*
Forward Totalized Mass Flow
[
+ [F3]
=
-MASS*
Reverse Totalized Mass Flow
[
+ [F4]
=
POWER**
Energy Flow Power
[
+ [
+ [F1]
=
+ENRG**
Forward Energy Flow
[
+ [
+ [F2]
=
-ENRG**
Reverse Energy Flow
[
+ [
+ [F3]
=
DIAG
Diagnostic
* Available only if “Mass Flow (static)?” = YES
** Available only if “Energy Option?” = YES
Содержание DigitalFlow DF868
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 10: ...Contents viii DigitalFlow DF868 Programming Manual 2 Channel...
Страница 104: ...Appendix A Menu Maps 96 DigitalFlow DF868 Multipurpose Ultrasonic Liquid Flowmeter Programming Manual 2 Channel...
Страница 118: ...Appendix B Data Records 110 DigitalFlow DF868 Multipurpose Ultrasonic Liquid Flowmeter Programming Manual 2 Channel...
Страница 150: ...Appendix E OPC Server 142 DigitalFlow DF868 Multipurpose Ultrasonic Liquid Flowmeter Programming Manual 2 Channel...
Страница 158: ...2 DigitalFlow DF868 Multipurpose Ultrasonic Liquid Flowmeter Programming Manual 2 Channel...
Страница 159: ......