Chapter 2. Standard Installation (FTPA Buffers)
18
BWT™ System Installation Guide
2.5.2 Inserting the BWT Transducer
Please read the following steps completely before beginning the installation of the BWT transducer.
IMPORTANT:
After you install the BWT transducers, you must allow the epoxy to cure for eight hours.
Note:
In gas applications, the typical minimum pressure should be 5 barg (75 psig) for the transducers to receive
any signal. A pressure higher than this minimum should result in a better signal-to-noise ratio.
1.
As shown below, place a small droplet (2 mm in diameter) of 3M Epoxy or equivalent on the center of the
transducer face (the
picture below shows the transducer without the junction box).
2.
Screw the transducer/junction box into the fitting at the end of the FTPA Buffer. When you begin to feel some
resistance, stop turning the transducer and wait 5 seconds to give the couplant a chance to spread out.
IMPORTANT:
In the next step,
DO NOT
overtighten the transducer.
3.
Torque the transducer to a maximum of 15-18 ft-lb (20- 25 N-m).
4.
Verify that the junction box is properly oriented as discussed in the previous section (see
5.
To connect the transducer electrical cables, see the
Startup or User’s Manual that was supplied with your
flowmeter.
Содержание BWT System
Страница 1: ...panametrics com Flow 916 058 F July 2021 BWT System Installation Guide ...
Страница 2: ......
Страница 4: ...ii no content intended for this page ...
Страница 8: ...Contents BWT System Installation Guide ...
Страница 16: ...Chapter 1 General Information 8 BWT System Installation Guide no content intended for this page ...
Страница 52: ...Chapter 5 Specifications 44 BWT System Installation Guide no content intended for this page ...
Страница 54: ...46 BWT System Installation Guide no content intended for this page ...
Страница 57: ......