020-17467-00 Instructions For Use PadMixer 1000 Premium RevD
39
Figure 43: Printer with paper roll
If required, press the FEED button to feed the paper in the printer or the PRINT button to print
out a test label.
Figure 44: Print and Feed buttons of the printer
6.5 Adjusting the inflation pressure
The outgoing pressure (to the biocontainer) can be adjusted by means of the manometer
underneath the mixing head cover plate (an L-shaped plate). This pressure is pre-set to a
maximum of 500 mbar.
If however a higher air flow rate is required, this pressure can be adjusted. There is no risk
for creating an overpressure inside the bag since the bag detections will stop the inflation in
time.
Note however that a too high gas flow might cause issues if certain filters are present on the
gas inlet tubing, or if the gas inlet tubing has a very small ID or shows locally diameter
reductions.
To change the pressure, the inflation must be switched on and the outlet tubing (towards the
bag) must be kept closed. Only then the manometer will indicate the outgoing pressure on
the display.