5
Operation and settings
42
5200-en_V2.0_06/20
5.2.2
Data screen – current measured values
The
data
screen shows the following data:
• The measured fine dust values of the individual fractions
PM
• The values measured by the weather station
• Particle number concentration
Cn
All of the data is saved every minute.
Menu path
Main menu > data
Data screen
Designation
Measured value
PM1
Fine dust values of different frac-
tions.
Display of sliding average (averaged
over 900 seconds*).
Update: every 30 seconds.
PM2.5
PM4
PM10
PMtot
Cn
Particle number concentration.
Update: every 3 seconds.
humidity
Relative humidity.
Update: every 2 minutes.
temp
Air temperature.
Update: every 2 minutes.
pressure
Air pressure.
Update: every 2 minutes.
* Value specified for operation as certified measuring equipment.