5
Operation and settings
5.1
Turning the device on
Turning on the power switch on the back of the controller starts the operating
system and then the startup manager. After the license is verified, the
firmware
Fidas
starts.
The following things occur automatically:
• The text
NaN
appears briefly on the touch screen.
• The measuring process begins. The data is automatically stored in the
internal memory. It takes about two minutes before the first PM values
are displayed.
• The main firmware menu appears.
The warm-up time, from switching on the device until valid measured data is
available, is about 10 to 15 minutes.
Starting firmware from operating system
If the firmware has been closed but the operating system is still running, the
firmware can be started by double-clicking the file
startupmanager.exe
(
Desktop > Startup
).