5.12. System
Doser management
group box
Default settings
Program doser with default settings. We suggest use this button if
your doser after firmware upgrade not work properly.
Clean memory
Erase doser memory. Should be only performed after Pacific Sun
Service request.
Soft reset
Generate reset signal for doser.
Emergency
connect
Should be performed if software can’t connect with doser due
firmware incompatibility. After connection firmware update (with
proper corresponding to software firmware) should be performed.
Memory backup
Generate .mbf file(memory backup file). This file can be send to
our service for diagnose (if something not work properly).
Restore backup
Allow import .mbf file to doser memory.
Calibration backup
Generate file with pump calibration config.
Restore calibration
Allow import calibration backup file to doser memory.
Sensor check
When enabled (check-box) on LCD screen show floating level
sensor status. Can be used for diagnose proper switch connection/
readings.