Without atria events within certain periods, without ventricular activity during certain
periods, the pacemaker delivers a pacing stimulus to the ventricle at the end of the
programmed
A-V INTERVAL.
The pacing rate can be limited by the
UPPER RATE
SO that the pacemaker does not pace the ventricle too fast in the presence of atria
arrhythmias. When pacing in the VDD mode, the pacemaker will pace the ventricle at the
programmed base rate
and with the programmed A-V interval in the absence of detected
atria or ventricular intrinsic events.
8.
DEVICE MAINTENANCE
8.1.
Cleaning
Zeus-1 Temporary Pacemaker
The Model Zeus-1 temporary pacemaker can be cleaned using a sponge or cloth
moistened with water or 70% isopropyl alcohol.
Note: Do not expose the unit to ethers, acetone, or chlorinated solvents as these
may damage the case or labels.
Caution: The Model Zeus-1 must not be immersed in water or cleaning
agents.
Severe damage to the device may occur.
Connecting Cables
The Patient Cables supplied are sterilized by ethylene oxide.
Prior to sterilization, Patient Cable should be cleaned thoroughly with a mild
detergent or 70% isopropyl alcohol to remove all visible blood and body fluids.
The cables may be immersed for cleaning. The cables must be thoroughly dried
after cleaning. Inspection and testing by a qualified technician should be done
after cleaning to verify proper cable function.
8.2.
Safety & Technical checks
Safety and technical checks should be carried out
on the
Model Zeus-1 at least
once every 6 months and after any malfunction or accident. Pacetronix does not
recommend field repair of the device. For service or repair contact your local
Pacetronix representative at the appropriate address.
9
WARNINGS