As an additional safety precaution the battery and contact terminals
should be checked for corrosion at regular intervals. Use of other than
the recommended batteries may result in one of the following
conditions:
(1) less than 24 hours of operation after the low-battery indicator
comes on,
(2) degraded pacemaker performance, and/or (3) overall reduced
battery life.
7.
MODE DEFINITIONS
Single Chamber Modes
The Model Zeus-1 will pace in four single chamber modes: AOO, VOO, AAI, and VVI.
AOO and VOO Modes
The asynchronous mode (A00 and VOO) emit a pacing pulse into one chamber, either the
atrium (AOO) or the ventricle (VOO),at the end of the base rate interval, regardless of
any intrinsic activity. When pacing in the A00 mode, the pacemaker will pace the atrium
asynchronously at the programmed base Rate
.
The atria and ventricular channels are
continuously blanked. When pacing in the VOO mode, the pacemaker will pace the
ventricle asynchronously at the programmed base Rate
.
AAI and VVI Modes
In the inhibited (or demand) modes (AAI and VVI), when an intrinsic beat is sensed
outside of a refractory period and before timeout of the base rate interval, the pacemaker
inhibits the output pulse. If no intrinsic beat is sensed before the timeout of the base rate,