OPERATOR’S    Oxi-Go™ Premier 

MANUAL    

With 2 Way Display

 

VER2.0C16 

Fingertip Pulse Oximeter

 

General Description   

Haemoglobin Saturation is percentage of HbO2 (Oxyhemoglobin), compounded with 
oxygen, in haemoglobin (Hb). In other words, it indicates the saturation of HbO2 in blood. 
It is a crucial physiological parameter for Respiration System. Many respiratory diseases 
can lead to decrease of SpO

2

 saturation. Moreover, such factors as organic automatic 

adjustment malfunction derived from anesthesia, side effect of major operation and 
injures aroused from physical check can also cause problems about SPO2 saturation 
variation, which contributed to such adverse symptoms such as faint, vomit and feel 
weakness and so on. What’s more, if can not get the right remedy, it can lead to dangers 
to human’s life. Hence, it is very feasible for doctors to analyze patients’ SPO2 
saturation as reference to detect the root of the causes, which is more effective to 
relieve patients’ pain and conduct suitable therapy plan as soon as possible.   

The fingertip pulse oximeter features of small dimension, low power consumption, 
easy-to-use and convenience  for carry. As long as you put your finger  in the device 
which read data though a light-sensitive probe, then the associated reading will be 
displayed on the indicator. It is proved by clinical trial Pulse Oximeter have outstanding 
precise and could repeat to measure the SPO2 more accurately and stably. 

 

 

Measurement Principle 

The fingertip pulse oximeter contains a dual light source and photodetector.Bone, 
tissure, pigmentation, and venous vessels normally absorb a constant amount of light 
over time. The arteriolar bed normally pulsates and absorbs variable amounts of light 
during systole and diastole, as blood volume increases and decreases. The ratio of light 
absorbed at systole and diastole is translated into an oxygen saturation 
measurement.This measurement is referred to as SpO

2

The principle of the oximeter rely on the property of spectrum which collect data from 
hemoglobin and oxyhemoglobin in glow area and area of approximate infrared source 
and apply the formular “Lamnert Beer” to establish the evaluating equation. In fact, the 
basic calculating technique is to combine electro-photo SPO2 detecting technology with 
recording mock pulse rate technology to get the result, the process demonstrate as 
below:    

Firstly, use glow with wavelength of 660cm and approximate infrared with wavelength of 
940cm to put light on the finger tip ready to read analogue data. 

Secondly, to process the data by electric circuit and microprocessor, then convey the 
data to LED display for reading. 

   

Diagram of Operation Principle 

1.  Red and Infrared-ray Emission Tube 
2.  Red and Infrared-ray Receipt Tube 

 

Technical Specifications 

1.    Display mode: LED. 

Measuring range for SpO

2

: 70%-99%   

Measuring range for PR: 30-235 BPM   

    PR display mode: bargraph.       

    Low power indication: 

 

2.    Battery standard: Two AAA 1.5V Alkaline Battery 

3.    Power consumption: Less than 40mA 

4.    Resolution: ±1% for SPO2 and ±1BPM for Pulse Rate 

5.    Measurement accuracy:   

SpO2: 80%--99%, ±2%; 70%--79%, ±3%; 

≤69%, no definition.   

PR: 30-99BPM, ±2 BPM; 100-235 BPM ±2%   

Operation Temperature: 41°F - 104°F, Storage Temperature: -4°F - 131°F

 

    Ambient Humidity: 

80% in operation, 

93% in storage   

 

6.    Sensitive degree test: Under the condition of poor perfusion, when the amplitude of 

pulse waveform is 0.6% detected by BIO-TEK INDEX tester. Then it can be proved 
the sensitive degree is Ok. 

7.    Resistance capacity against ambient light: Device work normally when mixed noise 

produced by BIO-TEK INDEX Pulse Oximeter tester   

8.    Automatic power off: When no finger in the device for  8 seconds, it will power off 

automatically. 

 

Precautions for use 

Do not use the pulse oximeter in an MRI or CT environment 

Do not use the pulse oximeter in situations where alarms are required. The 
device has no alarms. 

Explosion hazard: 

Do not use the pulse oximeter in an explosive atmosphere. 

The pulse oximeter is intended only as an adjunct in patient assessment. It must 
be used in conjunction with other methods of assessing clinical signs and 
symptoms. 

Check the pulse oximeter sensor application site frequently to determine the 
positioning of the sensor and circulation and skin sensitivity of the patient.   

Do not stretch the adhesive tape while applying the pulse oximeter sensor. This 
may cause inaccurate readings or skin blisters. 

Before use, carefully read the manual. 

The pulse oximeter has no SpO2 alarms; it is not for continuous monitoring, as 
indicated by the symbol.   

Prolonged use or the patient’s condition may require changing the sensor site 
periodically. Change sensor site and check skin integrity, circulatory status, and 
correct alignment at least every 4 hours. 

Inaccurate measurements may be caused by: 

10 

Autoclaving, ethylene oxide sterilizing, or immersing the sensors in liquid may 
cause inaccurate readings. 

11 

Significant levels of dysfunctional hemoglobins(such as carbonxy-hemoglobin or 
methemoglobin) 

12 

Intravascular dyes such as indocyanine green or methylene blue 

13 

SpO2 measurements may be adversely affected in the presence of high ambient 
light. Shield the sensor area (with a surgical towel, or direct sunlight, for example) 
if necessary. 

14 

Excessive patient movement 

15 

High-Frequency electrosurgical interference and defibrillators 

16 

Venous pulsations 

17 

Placement of a sensor on an extremity with a blood pressure cuff, arterial 
catheter, or intravascular line 

18 

The patient has hypotension, severe vasoconstriction, severe anemia, or 
hypothermia 

19 

The patient is in cardiac arrest or is in shock 

20 

Fingernail polish or false fingernails may cause inaccurate SpO2 readings. 

21 

WARNING!

  Oximeters are not for children’s play.    Keep the oximeter away 

from young children.    Small parts such as the battery door, batteries etc. may 
be hazardous if swallowed. The lanyard may cause strangulation in conditions 
that may cause it to twist around the neck.

 

Follow  local ordinances and recycling instructions regarding disposal or 
recycling of the device and device components, including batteries.

 

Product Properties 

1.    Easy to use.   

2.  Small dimension, light in weight (total weight is about 50g including batteries) and 

convenient for portable. 

3.    Lower power consumption;  originally-equipped two AAA batteries can last to  30 
hours.   

4.    Low voltage warning will be indicated in visual window when battery voltage is too 

low and normal application of the Oximeter might be influenced. 

5.    No signal can be tested, device will power off automatically in 8 seconds.   

Product Operation Scope 

The fingertip oximeter can be used to measure human Haemoglobin Saturation and 
heart rate through finger. The product is suitable for use in family, hospital (including 
clinical use in internist/surgery, Anaesthesia, paediatrics, intensive care and etc.) 
Oxygen  Club, social medical organizations, physical care in sports (It can be used 
before or after sports. Operation in sport procedure is not recommended) and etc. 
The product is not suitable to monitor patient 

continuously. Spot Monitoring only.

 

The pulse oximeter requires no routine calibration or maintenance other than 
replacement of batteries.

 

Operation Instructions 

1.    Installing two AAA batteries into battery cassette before closing its cover. 

2.    Open the clamp shown as in the picture below: 

3.    Put one of your fingers into rubber hole of the Oximeter (it is better to let your finger 

touch the bottom.) before releasing the clamp 

4.    Press the switch button for one time on front panel. 

5.    Do not rock your finger when starting test. Recommend you do not move your body 

at the same time.   

6.    Read correspondent data from display screen. 

Declaration:

  Please use the medical alcohol to clean the rubber touching the finger 

inside of Oximeter, and clean the test sensor using alcohol before and after operation. 
(The rubber inside of the Oximeter is medical rubber, which has no toxin, and no harmful 
to the skin of human being).             

When you put finger into the Oximeter, your nail surface must be 
upward. 