3.6.2.4
System Link Configuration Table
NAME FUNCTION NOTES
LK1
ANALOGUE 0V TO CHASSIS
LK2
DIGITAL 0V TO ANALOGUE 0V
LK3
NON – CONTROLLER CRYO ENABLE
LK4
HEATER SNAP SWITCH BYPASS
LK5
FIT IF NO OEM CONTROLLER
LK6 A/B
LK6A = NON PM140 ENDPOINT
LK6B = PM140 ENDPOINT
SEE LK7 A/B
LK7 A/B
LK7A = NON PM140 ENDPOINT
LK7B = PM140 ENDPOINT
SEE LK 6 A/B
LK8
+24V DC TO RL2 COM1
LK9
+24V DC TO RL2 COM2
LK10
+24V DC TO RL6 COM1
LK11
+24V DC TO RL6 COM2
LK12
+24V DC TO RL7 COM1
LK13
+24V DC TO RL7 COM2
LK14 A/B
LK14A = NON DRY PUMP FITTED
LK14B = DRY PUMP FITTED
LK15
PUMP CURRENT BYPASS
LK16
+24V DC TO RL8 COM2
LK17
NON – CONTROLLER HEATER ALARM
LK18
NON – CONTROLLER HEATER ENABLE
LK19
SPARE INT/LOCK 1 BYPASS
NOT FITTED 100/133
LK20
SPARE INT/LOCK 2 BYPASS
LK21 A/B
LK21A = PURGE SWITCH NOT FITTED
LK21B = PURGE SWITCH FITTED
Table 3-4: System link configuration
3.6.2.5
System LED Monitoring Table
NAME COLOUR
MONITORING
LED1 GREEN
+24V
DC
LED2 RED +15V
DC
LED3 YELLOW
-15V
DC
LED4 GREEN
+5V
DC
LED5
RED
CM COMP OK
LED6
RED
PUMP CURRENT SWITCH
LED7
RED
N2 PRESSURE SWITCH
LED8 RED WATER
ONE
LED9 RED WATER
TWO
LED10 RED WATER
THREE
LED11 RED VAC
STAT
LED12
RED
RL10 (HOIST CONTROL)
LED13 RED HOIST
LED14
RED
SPARE INTERLOCK 1
LED15
RED
SPARE INTERLOCK 2
LED16
RED
PUMP PURGE SWITCH
LED17
RED
RL15 (MASTER/SLAVE) ACTIVE SLAVE
Table 3-5: System LED monitoring
Note that when the interlock chain is complete, all LEDs are illuminated.