Cypher SPM
User Guide
5. Remember to cut off extra tubing before connecting gas lines to the cell body.
l
If you are using a Cypher ES that shipped before 2014
it is likely that you need to cut off the extra length of
tubing that extends past the end of the ferrule on
the gas lines.
l
Cut the tubing so that is flush with the ferrule. You
only need to do this for the end of the line that con-
nects to the cell body.
l
Failure to cut off the tubing will not break anything,
but it will make it much more difficult to get a good
seal between the gas line and the cell body.
6. When unlocking the sample stage from the scanner you only need three turns.
l
To unlock the sample stage from the scanner, turn
the locking screw counter-clockwise three turns.
l
Turning the locking screw more than three turns may
cause the screw to come all the way out.
l
If the screw comes all the way out, you may need to
lean the scanner forward until the screw slides out to
a place where you can put it back on the wrench.
7. Don't apply more than 200 mbar pressure to the cell.
l
When checking for leaks in the cell, it is useful to
apply pressure to the chamber by using a syringe.
l
Pressures above 200 mbar might damage the dia-
phragm.
10.8 Disassembling the sample stage from the cell body
Separating the sample stage and cell body is the same procedure regardless of which cell body and
sample stage you are using. Your cell body and sample stage may differ, but the way the two
attach is universal. For example, the following image shows an ambient stage and fluid cell body
combination on the left and a heater stage and gas cell body combination on the right.
Page 79 of 243
10 Cell Body and Sample Stage Guide
Содержание Cypher SPM
Страница 1: ...Cypher SPM User Guide Part No XX XXXX XXX...
Страница 8: ...Part I Getting Started System safety and overview...
Страница 165: ...Cypher SPM User Guide Sub Figure a First view of new cantilever 24 Engaging the cantilever on the sample Page 165 of 243...
Страница 167: ...Cypher SPM User Guide Sub Figure c Cantilever in focus See 9 24 Engaging the cantilever on the sample Page 167 of 243...
Страница 169: ...Cypher SPM User Guide Sub Figure a Cantilever Zoom See 10 24 Engaging the cantilever on the sample Page 169 of 243...
Страница 170: ...Cypher SPM User Guide Sub Figure b Laser Spot ON See 11 Page 170 of 243 24 Engaging the cantilever on the sample...
Страница 173: ...Cypher SPM User Guide Sub Figure a Offset 24 Engaging the cantilever on the sample Page 173 of 243...
Страница 176: ...Cypher SPM User Guide Sub Figure a Master Panel Main Tab Page 176 of 243 24 Engaging the cantilever on the sample...
Страница 186: ...Cypher SPM User Guide Sub Figure a Before the Move Page 186 of 243 25 AC mode imaging...
Страница 189: ...Cypher SPM User Guide 26 26 Contact Mode Imaging In Air 26 Contact mode imaging in air Page 189 of 243...
Страница 216: ...Part VI Shipping and Maintenance This part of the user guide covers moving and shipping the instrument...