Cypher SPM
User Guide
The benefits of such a direct drive mechanism are similar to those of magnetic actuation (iDrive),
but with the advantage of being applicable to cantilevers of all shapes and sizes. Photothermal
excitation allows quantitative imaging in all environments, and the benefits are especially note-
worthy in liquids, where blueDrive enables stable imaging for hours. The only drawback is that can-
tilevers may require a gold coating to achieve high amplitudes with photothermal excitation.
More details about the benefits of blueDrive are described in the following webinar:
AFM Imaging and Nanomechanics with New blueDrive™ Photothermal Excitation
”.
16.2 How much light power do you need?
Depending on the cantilever shape, the coating, or the imaging environment, up to 10 mW of blue
light might be necessary to drive the cantilever substantially, or as little as 0.1 mW might be
enough.
As shown in the following image, blueDrive is supplied with five filter cubes. The purpose of the fil-
ter cubes is to attenuate the blue light power to a desired level.
Choosing the correct filter cube ensures that the right amount of blue light power reaches the can-
tilever in order to avoid having excess blue light hitting the cantilever. This minimizes cantilever
heating and maximizes the performance of blueDrive.
Page 125 of 243
16 blueDrive photothermal excitation
Содержание Cypher SPM
Страница 1: ...Cypher SPM User Guide Part No XX XXXX XXX...
Страница 8: ...Part I Getting Started System safety and overview...
Страница 165: ...Cypher SPM User Guide Sub Figure a First view of new cantilever 24 Engaging the cantilever on the sample Page 165 of 243...
Страница 167: ...Cypher SPM User Guide Sub Figure c Cantilever in focus See 9 24 Engaging the cantilever on the sample Page 167 of 243...
Страница 169: ...Cypher SPM User Guide Sub Figure a Cantilever Zoom See 10 24 Engaging the cantilever on the sample Page 169 of 243...
Страница 170: ...Cypher SPM User Guide Sub Figure b Laser Spot ON See 11 Page 170 of 243 24 Engaging the cantilever on the sample...
Страница 173: ...Cypher SPM User Guide Sub Figure a Offset 24 Engaging the cantilever on the sample Page 173 of 243...
Страница 176: ...Cypher SPM User Guide Sub Figure a Master Panel Main Tab Page 176 of 243 24 Engaging the cantilever on the sample...
Страница 186: ...Cypher SPM User Guide Sub Figure a Before the Move Page 186 of 243 25 AC mode imaging...
Страница 189: ...Cypher SPM User Guide 26 26 Contact Mode Imaging In Air 26 Contact mode imaging in air Page 189 of 243...
Страница 216: ...Part VI Shipping and Maintenance This part of the user guide covers moving and shipping the instrument...