Cypher SPM
User Guide
due to subtle position shifts in the relationship of the optical components, the light may settle in a
new location. Once the light tube is installed and any corrections are made to the calibration, the
spot accuracy of the Spot On routine is very repeatable. If the Spot On software routine appears to
misalign the light on the back of the cantilever, it may be for the following reasons:
l
The light tube you installed may be clamped in a different position from when it was tested at
the factory. With the light tube powered on, loosen and jiggle the light tube to allow it to re-
seat in the head. Re-tighten the clamp. If the light moves back to the same (unwanted) location
repeatedly then consider recalibration.
l
If the light tube was purchased at a later time from when the system was delivered or it was
exchanged due to a repair then the Spot On parameters were calibrated on a different instru-
ment and small system to system differences make the Spot On calibration parameters in the
light tube incorrect for your system.
15.3.1 Readjusting the Spot On calibration
The Spot On software routine relies on two parameters stored in the light tube calibration inform-
ation block (info block). These parameters are set at the factory but may need to be changed if the
light tube is exchanged.
Before you decide to adjust the Spot On calibration, consider the following points:
l
Subtle changes in the relative position of the optical emitter and focusing lenses inside the light
tube can occur during warm up after installation. You might find that the error in the Spot On
positioning reduces after warm up.
l
If the instrument is left powered off for an extended amount of time, for example, overnight or
for a day, you may find that the Spot On position may be off. Allow ample time for the system
to reach thermal equilibrium before deciding to make adjustments.
If you find that the Spot On position is still off after re-seating the light tube and allowing it to warm
up, perform the following procedure.
NEED THE NEW INTERFACE PATHS FOR SOFTWARE STEPS.
1. Allow an hour for the light tube to come to thermal equilibrium with the system.
2. Select Options > Show Reset SpotOn Menu Item.
Page 119 of 243
15 Light tubes
Содержание Cypher SPM
Страница 1: ...Cypher SPM User Guide Part No XX XXXX XXX...
Страница 8: ...Part I Getting Started System safety and overview...
Страница 165: ...Cypher SPM User Guide Sub Figure a First view of new cantilever 24 Engaging the cantilever on the sample Page 165 of 243...
Страница 167: ...Cypher SPM User Guide Sub Figure c Cantilever in focus See 9 24 Engaging the cantilever on the sample Page 167 of 243...
Страница 169: ...Cypher SPM User Guide Sub Figure a Cantilever Zoom See 10 24 Engaging the cantilever on the sample Page 169 of 243...
Страница 170: ...Cypher SPM User Guide Sub Figure b Laser Spot ON See 11 Page 170 of 243 24 Engaging the cantilever on the sample...
Страница 173: ...Cypher SPM User Guide Sub Figure a Offset 24 Engaging the cantilever on the sample Page 173 of 243...
Страница 176: ...Cypher SPM User Guide Sub Figure a Master Panel Main Tab Page 176 of 243 24 Engaging the cantilever on the sample...
Страница 186: ...Cypher SPM User Guide Sub Figure a Before the Move Page 186 of 243 25 AC mode imaging...
Страница 189: ...Cypher SPM User Guide 26 26 Contact Mode Imaging In Air 26 Contact mode imaging in air Page 189 of 243...
Страница 216: ...Part VI Shipping and Maintenance This part of the user guide covers moving and shipping the instrument...