Cypher SPM
User Guide
9. Close the valve.
l
Make sure the valve is in the off position, if it was not
already.
l
The cell is now fully sealed on all sides.
10. Take the assembled cell out of the glovebox.
l
Place the cell and everything attached into the load
lock and close it.
l
DO NOT EVACUATE. There is no need for evacuation
when taking things back into the ambient atmo-
sphere.
l
Open the door on the ambient side and take the cell
to the AFM.
Tip:
Some customers like to first place their cell and
tubing inside of a jar, like the dessicator shown on
the right. This gives you even more time and peace
of mind when transporting the cell back to the
AFM.
12.4 Mount the sealed cell onto the AFM
The faster you work through the following steps, the smaller the chance of impurities diffusing into
the sealed cell. The membrane is made of Kalrez and is quite impervious to water and oxygen.
Nevertheless, it is best to get pure dry gas flowing through the cell as quickly as possible.
12 Glovebox Protocols
Page 102 of 243
Содержание Cypher SPM
Страница 1: ...Cypher SPM User Guide Part No XX XXXX XXX...
Страница 8: ...Part I Getting Started System safety and overview...
Страница 165: ...Cypher SPM User Guide Sub Figure a First view of new cantilever 24 Engaging the cantilever on the sample Page 165 of 243...
Страница 167: ...Cypher SPM User Guide Sub Figure c Cantilever in focus See 9 24 Engaging the cantilever on the sample Page 167 of 243...
Страница 169: ...Cypher SPM User Guide Sub Figure a Cantilever Zoom See 10 24 Engaging the cantilever on the sample Page 169 of 243...
Страница 170: ...Cypher SPM User Guide Sub Figure b Laser Spot ON See 11 Page 170 of 243 24 Engaging the cantilever on the sample...
Страница 173: ...Cypher SPM User Guide Sub Figure a Offset 24 Engaging the cantilever on the sample Page 173 of 243...
Страница 176: ...Cypher SPM User Guide Sub Figure a Master Panel Main Tab Page 176 of 243 24 Engaging the cantilever on the sample...
Страница 186: ...Cypher SPM User Guide Sub Figure a Before the Move Page 186 of 243 25 AC mode imaging...
Страница 189: ...Cypher SPM User Guide 26 26 Contact Mode Imaging In Air 26 Contact mode imaging in air Page 189 of 243...
Страница 216: ...Part VI Shipping and Maintenance This part of the user guide covers moving and shipping the instrument...