4.Advanced User Guidebook
43
How to Measure with Cursors
Push the
Cursor
button to turn cursors on and display the cursor menu. Push it
again to turn cursors off.
The Cursor Measurement for normal mode:
The description of the
cursor menu
is shown as the following table:
Function
Menu
Setting
Description
Type
Voltage
Time
Time&Voltag
e
AutoCursr
Display the voltage measurement cursor and
menu.
Display the time measurement cursor and
menu.
Display the time and voltage measurement
cursor and menu.
The horizontal cursors are set as the
intersections of the vertical cursors and the
waveform
Line
Type
(Time&Volt
age type)
Time
Voltage
Makes the vertical cursors active.
Makes the horizontal cursors active.
Window
(Wave
zoom
mode)
Main
Extension
Measure in the main window.
Measure in the extension window.
Line
a
b
ab
Turn the
M
knob to move line a.
Turn the
M
knob to move line b.
Two cursors are linked. Turn the
M
knob to
move the pair of cursors.
Source
CH1
CH2
CH3
CH4
Display the channel to which the cursor
measurement will be applied.
Perform the following operation steps for the time and voltage cursor
measurement of the channel CH1:
1. Push
Cursor
to display the cursor menu.
2. In the right menu, select
Source
as
CH1
.
3. Press the first menu item in the right menu, select
Time&Voltage
for Type,
two blue dotted lines displayed along the horizontal direction of the screen,
two blue dotted lines displayed along the vertical direction of the screen.
Cursor measure window at the left bottom of the screen shows the cursor
readout.
4. In the right menu, select
Line Type
as
Time
to make the vertical cursors
active. If the
Line
in the right menu is select as
a
, turn the
M
knob to move