4.Advanced User Guidebook
61
The description of the
Acqu Mode
menu is shown as follows:
Function Menu
Setting
Description
Acqu
Mode
Sample
Normal sampling mode.
Peak Detect
Use to capture maximal and minimal
samples. Finding highest and lowest points
over adjacent intervals. It is used for the
detection of the jamming burr and the
possibility of reducing the confusion.
Average
4, 16, 64, 128
It is used to reduce the random and don't-care
noises, with the optional number of averages.
The description of the
Record Length
menu is shown as follows:
Function Menu
Setting
Description
Length
1000
Choose the record length
(For
XDS3102AP/XDS3202A,
40M
is
available on both single CH and double CH)
10K
100K
1M
10M
20M
40M (Single CH)
The description of
PERF Mode
menu (only for XDS3000-A series) is shown as
follows:
Function Menu
Setting
Description
PERF Mode
8-bit
Set the vertical resolution (A/D)
12-bit
14-bit
(only for
XDS3102AP/XDS3202A)
This menu is only for XDS3000-A series.
When the sample rate <=250MS/s, ADC resolution is set to 12-bit by default;
When the sample rate >250MS/s, and both channels are turned on, ADC resolution is
set to 8-bit by default;
When the sample rate <250MS/s, and only one channel is turned on, ADC resolution
can be set to 8-bit or 12-bit in this menu.
Note
: For
XDS3102AP/XDS3202A, the selection of 8-bit, 12-bit or 14-bit is not affected by
sample rate.
The description of the
Intrpl
menu is shown as follows:
Function Menu
Setting
Description
Intrpl
Sinx/x
Use sine(x)/x interpolation
x
Use linear interpolation
Содержание Dual-Channel Series
Страница 1: ...XDS3000 Dual Channel Series Digital Storage Oscilloscopes User Manual www owon com ...
Страница 29: ...3 Junior User Guidebook 22 Double Zoom Single Zoom ...
Страница 42: ...4 Advanced User Guidebook 35 Figure 4 2 Rectangle window Figure 4 3 Blackman window ...
Страница 43: ...4 Advanced User Guidebook 36 Figure 4 4 Hanning window Figure 4 5 Kaiser window ...