User Manual
– Safety information
NIMXEN020I
Owandy Radiology SAS
13
3.3.1
Electromagnetic emissions
In accordance with the IEC 60601-1-2 Ed4 standard, I-MAX 3D is suitable for use in the
electromagnetic environment specified below.
The customer or user of the system must ensure that it is used in the said environment.
Emissions test
Compliance
Electromagnetic environment
RF emissions
CISPR 11
Group I
I-MAX 3D uses RF energy only for its internal
function. Therefore, its R.F. emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
Class A
I-MAX 3D is suitable for use in all establishments
other than domestic and those directly connected
to the public low voltage power supply network
that supplies buildings used for domestic
purposes.
Harmonics emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Содержание i-max touch 3D
Страница 1: ...EN USER MANUAL I MAX 3D NIMXEN020I June 2022...
Страница 2: ......
Страница 5: ...User Manual Revision history NIMXEN020I Owandy Radiology SAS THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 6: ......
Страница 119: ...User Manual NIMXEN020G Owandy Radiology SAS 109...