![Ovizio Imaging Systems ILine F Pro Скачать руководство пользователя страница 59](http://html1.mh-extra.com/html/ovizio-imaging-systems/iline-f-pro/iline-f-pro_user-manual_1663646059.webp)
59
Hardware
Camera error
Description:
OsOne informs the user that a Camera error has been detected.
Cause(s):
•
OsOne could not connect with the camera
•
There was an error during the transfer of the images
When?
This alarm can be triggered anytime
Type:
ERROR
Does it prevent from starting a monitoring?
YES
Does it interrupt the monitoring?
YES
Is it handled by OsOne?
YES, if this error
occurs during a monitoring. In that case the “Restart tool” will:
•
Close OsOne
•
Restart OsOne
•
Restart the monitoring
If the error is not automatically resolved a
fter 3 consecutive tries of the “Restart
tool”, the monitoring could
not continue because OsOne could not record any images.
How to fix it?
If this error occurs outside of a monitoring, it is recommended to restart OsOne.
If restarting OsOne did not solve the issue, please try to power off the microscope, wait a minute and power
the microscope on.
If the issue still occurs after having power off/on the microscope, please contact the Ovizio support
([email protected]).
How to prevent it?
There is no way to prevent a camera error.
Содержание ILine F Pro
Страница 1: ...1 Ovizio Imaging Systems ILine F Pro User Manual...
Страница 12: ...12 Introduction...
Страница 16: ...16 Connect...
Страница 17: ...17 Get Started Description BioConnect drawing...
Страница 25: ...25 Press here to restart OsOne...
Страница 26: ...26 Measure...
Страница 31: ...31 Detailed views Live image Click on the Live image to enlarge it The following screen will be displayed...
Страница 39: ...39 Now you can disconnect the BioConnect pump from the Microscope pump motor...
Страница 41: ...41 Troubleshooting...
Страница 68: ...68 Configure...
Страница 76: ...76...
Страница 77: ...77...
Страница 86: ...86 Remote Access...
Страница 88: ...88 Maintenance and Servicing...