Fluidic issue
–
Cells in background images
Description:
OsOne informs the user that a potential fluidic issue could have occurred during the monitoring.
Cause(s):
Objects have been detected in the background images. The presence of objects in the background images
indicates that these objects are stuck in the flow cell because normally, the flush performed by OsOne during
the acquisition of the background images should have made them disappear.
When?
This alarm is triggered during the monitoring, when OsOne completed the computation of the last recorded
images.
Type:
ERROR
Does it prevent from starting a monitoring?
NO
Does it interrupt the monitoring?
YES
Is it handled by OsOne?
YES. If a fluidic issue occurs during the monitoring, OsOne will automatically (automatic flush wizard):
•
Stop the monitoring
•
Start a flush sequence of 5 minutes
•
Restart the monitoring.
If this error occurs 5 times consecutively with
out being resolved, the above automatic actions won’t be
performed anymore.
Содержание ILine F Pro
Страница 1: ...1 Ovizio Imaging Systems ILine F Pro User Manual...
Страница 12: ...12 Introduction...
Страница 16: ...16 Connect...
Страница 17: ...17 Get Started Description BioConnect drawing...
Страница 25: ...25 Press here to restart OsOne...
Страница 26: ...26 Measure...
Страница 31: ...31 Detailed views Live image Click on the Live image to enlarge it The following screen will be displayed...
Страница 39: ...39 Now you can disconnect the BioConnect pump from the Microscope pump motor...
Страница 41: ...41 Troubleshooting...
Страница 68: ...68 Configure...
Страница 76: ...76...
Страница 77: ...77...
Страница 86: ...86 Remote Access...
Страница 88: ...88 Maintenance and Servicing...