Calibration Alarms
Calibration file missing
Description:
OsOne informs the user that the calibration file is missing.
Cause(s):
OsOne
did
not
find
a
file
named
“
CalibrationImage.h5
”
inside
the
“C:
\Users\<OsOne
user>\AppData\Roaming\OsOne\Calibration
” folder
When?
This alarm is triggered at the start of OsOne.
Type:
ERROR
Does it prevent from starting a monitoring?
YES
Does it interrupt the monitoring?
NO
Is it handled by OsOne?
YES, if the BioConnect cartridge is not inserted at the start of OsOne: in that case, OsOne will automatically take
a new background image.
NO, if the BioConnect cartridge is inserted inside the microscope at the start of OsOne.
How to fix it?
By removing the BioConnect cartridge from the microscope the time that OsOne can take a new background
image.
How to prevent it?
By
never
deleting
the
“
CalibrationImage.h5
”
file
inside
the
“C:
\Users\<OsOne
user>\AppData\Roaming\OsOne\Calibration
” folder
.
Содержание ILine F Pro
Страница 1: ...1 Ovizio Imaging Systems ILine F Pro User Manual...
Страница 12: ...12 Introduction...
Страница 16: ...16 Connect...
Страница 17: ...17 Get Started Description BioConnect drawing...
Страница 25: ...25 Press here to restart OsOne...
Страница 26: ...26 Measure...
Страница 31: ...31 Detailed views Live image Click on the Live image to enlarge it The following screen will be displayed...
Страница 39: ...39 Now you can disconnect the BioConnect pump from the Microscope pump motor...
Страница 41: ...41 Troubleshooting...
Страница 68: ...68 Configure...
Страница 76: ...76...
Страница 77: ...77...
Страница 86: ...86 Remote Access...
Страница 88: ...88 Maintenance and Servicing...