Various events may occur during the setup of a bioreactor run or during the run itself such as bad quality images
leading to measurement issues. In all cases the alarms or notifications are displayed with a banner on top of the
live screen.
The next sections describe each of them in detail.
BioConnect Alarms
Pump motor stalled
Description:
OsOne informs the user that the pump motor has stalled.
Cause(s):
•
Pinched tubing during the setup in the lab.
•
Tube welding not popped open
•
BioConnect assembly defect (internal tubing pinched).
•
Aggregates/Clogging present in the cell culture.
•
Unexpected pinch valve action (software bug).
•
Wrong pump profile setting.
•
The pump motor could be too old and not working as expected (not powerful anymore).
When?
This alarm can be triggered anytime (before or during a monitoring) when the pump is executing one of the
following sequences:
•
Priming
•
Flushing
•
Backup pumping (pumping mode when not in an active monitoring session)
Type:
ERROR
Does it prevent from starting a monitoring?
YES
Содержание ILine F Pro
Страница 1: ...1 Ovizio Imaging Systems ILine F Pro User Manual...
Страница 12: ...12 Introduction...
Страница 16: ...16 Connect...
Страница 17: ...17 Get Started Description BioConnect drawing...
Страница 25: ...25 Press here to restart OsOne...
Страница 26: ...26 Measure...
Страница 31: ...31 Detailed views Live image Click on the Live image to enlarge it The following screen will be displayed...
Страница 39: ...39 Now you can disconnect the BioConnect pump from the Microscope pump motor...
Страница 41: ...41 Troubleshooting...
Страница 68: ...68 Configure...
Страница 76: ...76...
Страница 77: ...77...
Страница 86: ...86 Remote Access...
Страница 88: ...88 Maintenance and Servicing...