Finishing the prosthesis
The 3S27=L/R44 Foam Cover can be used for finishing the pelvic socket prosthesis. Allow a 60 mm "compres
sion allowance" to minimise the protective foam cover’s effect on knee function. When determining the proper
cover length, increase the linear thigh dimension by 30 mm and the linear shin dimension by the same amount.
Use 710D4 Torque Wrench to finish the prosthesis and secure all set screws with 636K13 Loctite. Please observe
the following torques:
Screws
Torque
501F9=* Flat headed screws on the lamination plate
(with Torx socket)
25 Nm (Wrench/Bit TX 40)
Small cap screw of lamination adapter
2 Nm
Set screws of rotation adapter 4R57
10 Nm / Torque above the rotation adapter
All other set screws
15 Nm
Finally, verify once again that all adjustments are optimal by having the prosthesis wearer walk with the finished
prosthesis. If necessary, readjust the prosthesis to compensate for influences by the foam cover.
3.9 Maintenance
Replacing the expansion springs
To support the swing phase (flexion), the joint contains tension springs on the left and right (order no.: 4G430,
red). These springs store energy during extension of the Helix
3D
Hip Joint in the stance phase and return the energy
at the beginning of flexion. This results in significantly faster initiation of hip flexion and larger ground clearance
during the swing phase. Two additional normal tension springs (order no.: 4G430, red) and two stronger tension
springs (order no.: 4G430=2, grey) are included.
Worn or defective expansion springs can be replaced as follows:
1. First open the front cap.
2. Then replace the defective expansion springs with new ones.
INFORMATION
To remove the expansion springs, use a slotted screwdriver
without
sharp
edges!
Before installing the 4G430 or 4G430=2 expansion springs, the eyelets of
the expansion spring
must
be greased throughly with 633F30 special grease
(Fig. 1).
1
3. Close the cap and insert the attachment pins in the lower section.
3.10 Maintenance instructions
Please make your patients aware of the following safety instructions:
CAUTION
Risk of injury as a result of excessive strain.
Depending on the environment and usage conditions, the func
tionality of the hip joint may be impaired. To reduce the risk for the patient, the hip joint must not be used after
noticeable functional changes have occured.
Noticeable functional changes may include poor response, incomplete extension, diminished stance phase sta
bility, deteriorating swing phase behaviour, abnormal noises, etc.
Measure:
Have the prosthesis inspected by a prosthetist.
Содержание Helix3D 7E10
Страница 25: ...Helix Ottobock 25...
Страница 213: ...Helix Ottobock 213...
Страница 259: ...Helix Ottobock 259...
Страница 283: ...Helix Ottobock 283...
Страница 311: ...Helix Ottobock 311 Ottobock H Ottobock Ottobock H H H Helix3D H...
Страница 312: ...312 Ottobock Helix I Helix3D Ottobock I Otto Bock HealthCare I Ottobock I 2 2 1 Helix3D Helix3D 7E10 L 7E10 R...
Страница 314: ...314 Ottobock Helix 3 1 1 2 3 4 5 cm 4 5 6 cm 5 A B...
Страница 315: ...Helix Ottobock 315 3 2 I 5 2 y y 2 1 1 2 1 3 2 y y 2 4 5...
Страница 316: ...316 Ottobock Helix 3 2 1 Helix3D 1 7Z53 1 2 2 1 2 4G576 I 3 2 Helix3D 5 I 4G576 2 5 mm 3 4 5...
Страница 318: ...318 Ottobock Helix 3 4 H 1 2 L A S A R Assembly 3 8 2...
Страница 319: ...Helix Ottobock 319 30 mm 5 mm 35 mm 1 2 6 3 4 5 0 5 mm...
Страница 321: ...Helix Ottobock 321 3 5 L A S A R Posture 1 2 L A S A R Posture 10 mm 30 mm Helix3D 3 6 PU 3 8 3 7...
Страница 327: ...Helix Ottobock 327 I Ottobock I ISO 15032 2 3 I Helix3D Ottobock 4 Helix3D 1 2 Helix3D...
Страница 328: ...328 Ottobock Helix 5 5 1 R L 130 990 g 275 g 146 mm 2 3 100 kg 10 C 60 C 5 2 5 3 5 3 1 5 3 2 CE 2017 745 5 3 3...
Страница 329: ...Helix Ottobock 329...
Страница 354: ......
Страница 355: ......