CAUTION

Independent manipulation of Axon-Bus components

Injury due to faulty control or malfunction of the Axon-Bus prosthetic system.

Manipulations to the Axon-Bus prosthetic system other than the tasks described in these instructions for use
are not permitted.

Do not damage the battery or separate the connection cables between the battery packs. 

The  Axon-Bus  prosthetic  system  and  any  damaged  Axon-Bus  components  may  only  be  opened  and/or
repaired by certified Ottobock Myo-Service technicians.

CAUTION

Incorrect electrode settings

Injury due to faulty control or malfunction of the product.

The electrodes are to be placed on intact skin only and with as much electrode-skin contact as possible. In
the case of strong interference from electronic devices, the position of the electrodes should be checked and
changed if necessary. If the interference cannot be eliminated or if you do not achieve the expected results by
adjustment or selection of the appropriate control programme, please contact Ottobock Myo-Service.

Set the electrode gains as low as possible to reduce the risk of malfunctioning due to strong electromagnetic
radiation  (e.g.  visible  or  concealed  theft  prevention  systems  at  the  entrance/exit  of  stores,  metal  detect­
ors/body scanners for persons (e.g. in airports) or other sources of strong electromagnetic interference (e.g.
high-voltage  lines,  transmitters,  transformer  stations,  computer  tomographs,  magnetic  resonance  tomo­
graphs, etc.).

Allow the patient to rest during the adjustment of the electrodes; otherwise, muscle fatigue will lead to incon­
sistent results. As a result, the therapist will tend to set the electrode gains too high.

CAUTION

Risk of pinching in elbow joint flexion region

Injuries due to pinching of body parts.

Ensure that fingers and other body parts are not in this area when bending the elbow joint.

CAUTION

Manual unlocking of elbow lock under load

Injury by release of elbow lock under load.

Particular caution should be exercised when unlocking the elbow lock while lifting heavy loads.

Be careful when unlocking the lock under such conditions due to the possibility of injury.

CAUTION

Penetration of dirt and moisture

Injury due to faulty control or malfunction of the Axon-Bus prosthetic system.

Do not let foreign particles or liquids get into the prosthetic arm.

Do not expose the prosthetic arm, especially the elbow joint, to dripping or splashing water.

Wear the prosthetic arm and especially the elbow joint under suitably resistant clothing in the rain.

NOTICE

Coating, gluing or painting the product

Damage or fracture due to chemical processes.

The product must not be coated, glued or painted.

5 Scope of Delivery and Accessories

5.1 Scope of delivery

1 pc. 12K501 AxonArm Ergo

1 pc. Lamination protection cover (disc with arrow)

1 pc. Lamination protection cover (hollow ball)

1 pc. Cable lock

8

Scope of Delivery and Accessories

AxonArm Ergo 12K501=*

Содержание AxonArm Ergo 12K501

Страница 1: ...Instructions for use qualified personnel 3 AxonArm Ergo 12K501...

Страница 2: ...1 2 2 AxonArm Ergo 12K501...

Страница 3: ...6 5 12 Adjusting the Humeral Rotation Feature 6 6 14 Installing the Pull Cable 6 7 14 Checking the Symmetry 6 8 14 Handling 7 15 Charging the Battery 7 1 15 Charging Receptacle 7 2 15 Determining the...

Страница 4: ...The movements of the Axon Bus gripping component are becoming increasingly slower 10 2 4 19 Axon 10 3 19 4 AxonArm Ergo 12K501 Table of contents...

Страница 5: ...of the elbow joint The Automatic Forearm Balance provides for harmonious movement control free swing and reduction of energy consumption It stores the energy that is released when the arm is extended...

Страница 6: ...gs 3 2 Conditions of use The product is designed for use by unilateral or bilateral amputees The product is intended exclusively for use on adults The product was developed for everyday use and must n...

Страница 7: ...ed theft prevention systems at the entrance exit of stores metal detectors body scanners for persons e g in airports or other sources of strong magnetic and electric al interference e g high voltage l...

Страница 8: ...high voltage lines transmitters transformer stations computer tomographs magnetic resonance tomo graphs etc Allow the patient to rest during the adjustment of the electrodes otherwise muscle fatigue w...

Страница 9: ...hetic glove e g 8S500 R L AxonSkin Natural when 8E500 Michelangelo hand is used Additional MyoBock components such as suction socket electrode electrode cable pull in tube and tube valve for suction s...

Страница 10: ...long co contraction on channels 1 and 2 of the AxonMaster releases or locks the lock A short co contraction switches the gripping mode of the gripper A pulse on channel 3 of the AxonMaster releases or...

Страница 11: ...Reduction in mechanical strength due to thread misalignment The circuit board and the edge of the lamination ring must be parallel to each other Left picture The EasyPlug has been snapped in properly...

Страница 12: ...isted 7 Slide the strap clamp under the lamination ring 8 Screw the strap clamp in 6 5 Setting the Preflexion The level of AFB compensation flexion assistance can be adjusted and adapted to the weight...

Страница 13: ...e forearm or the upper arm socket due to released lock The Axon Bus prosthethic system must always be flexed for donning and doffing The Axon Bus prosthethic system must be stored in flexed condition...

Страница 14: ...rntable friction is easily adjusted by turning of the external adjustment screw using a coin or similar object 6 7 Installing the Pull Cable CAUTION Removing the pull cable Injury due to malfunction o...

Страница 15: ...ing Receptacle The charging receptacle Fig 1 item 2 has the following functions Contacts for charging the battery LED display to indicate the current battery charge level LED display for providing fee...

Страница 16: ...1x short 1x long Beep Pulsating Turn the Axon Bus pros thesis solution on Push the button for one to four seconds Turn the Axon Bus pros thesis solution off Push the button for more than one second C...

Страница 17: ...that may cause undesired operation This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to part 15 of the FCC Rules These limits are designed to pr...

Страница 18: ...ing General AxonArm Ergo Reference number 12K501 Dimensions mm Approx 345 mm x d55 mm to d80 mm arm extended Weight without Axon Bus terminal device g Approx 840 Forearm length as delivered mm 305 Min...

Страница 19: ...ds Probable cause The Axon Bus prosthetic system is overheated Remedy 1 Allow the Axon Bus prosthetic system to cool for a few minutes 2 Do not use the Axon Bus prosthetic system at high ambient tempe...

Страница 20: ...sion SB_2021 02 FM483 SB_210x297 Otto Bock Healthcare Products GmbH Brehmstra e 16 1110 Wien Austria T 43 1 523 37 86 F 43 1 523 22 64 info austria ottobock com www ottobock com Ottobock 647G790 0 en_...

Отзывы: