7 Handling

7.1 Charging the Battery

The  757B501  AxonEnergy  Integral  must  be  charged  prior  to  initial  use.  The  757B501  AxonEnergy  Integral  may

only

  be  charged  with  the  757L500  AxonCharge  Integral.  The  charging  process  can  only  be  performed  with  the

13E500 AxonMaster connected.
1)  Connect the charging plug of the 757L500 AxonCharge Integral to the charging receptacle.

The beeper sounds briefly 2x.

The Axon-Bus prosthethic system is deactivated and the charging process starts automatically. 

2)  To activate the Axon-Bus prosthethic system, disconnect the charging plug and switch on the Axon-Bus pros­

thethic system.

INFORMATION: The prosthesis cannot be used during the charging process.

The product is powered by an integrated high-quality Li-Ion battery with sufficient capacity to operate for one full
day of usual everyday activities. Turning the product off during longer periods of passive use (e.g. air or rail travel,
visit to a theatre or cinema, etc.) will make the battery last longer. We recommend charging the product once a day
when  used  by  the  patient  on  a  daily  basis.  Electronic  battery  management  provides  the  patient  with  information
about the battery charge level (battery management).
The  charging  unit  consists  of  the  charging  plug  and  the  power  cord.  The  battery  charger  has  an  input  voltage
range of 100–240 V and may be operated in a mains frequency range of 50–60 Hz. 

7.2 Charging Receptacle

The charging receptacle (Fig. 1, item 2) has the following functions:

Contacts for charging the battery

LED display to indicate the current battery charge level

LED display for providing feedback on operating states

Button to turn the Axon-Bus prosthetic system on and off, display the battery charge level and open the Axon-
Bus gripping component in an emergency

Beeper for providing feedback on operating states

Button for activating the Bluetooth function.

7.3 Determining the Battery Charge Level

The battery charge level can be queried at any time.
1. With  the  Axon-Bus  prosthetic  system  switched  on,  press  the  charging  receptacle  button  (Fig.  2,  arrow)  and

hold for less than one second.

2. The LED is illuminated. The colour of the LED indicates the current battery charge level.

Battery empty

Orange LED is illuminated

Battery 50% charged

Yellow LED is illuminated

Battery fully charged

Green LED is illuminated

7.4 Switching On the Axon-Bus Prosthetic System

Press the charging receptacle button and hold for one second (Fig. 2, arrow).

The beeper emits two short signals.

7.5 Switching Off the Axon-Bus Prosthetic System 

Press the charging receptacle button and hold for more than one second (Fig. 2, arrow).

The beeper emits one long signal.

7.6 Safety Shutoff

The purpose of the safety shutoff is to protect the battery; it is triggered in case of:

Excessively high or low temperature

Overvoltage or undervoltage

Short circuit.

After a short circuit the battery charger must be connected and then disconnected in order for the electronics to be
activated.

15

AxonArm Ergo 12K501=*

Handling

Содержание AxonArm Ergo 12K501

Страница 1: ...Instructions for use qualified personnel 3 AxonArm Ergo 12K501...

Страница 2: ...1 2 2 AxonArm Ergo 12K501...

Страница 3: ...6 5 12 Adjusting the Humeral Rotation Feature 6 6 14 Installing the Pull Cable 6 7 14 Checking the Symmetry 6 8 14 Handling 7 15 Charging the Battery 7 1 15 Charging Receptacle 7 2 15 Determining the...

Страница 4: ...The movements of the Axon Bus gripping component are becoming increasingly slower 10 2 4 19 Axon 10 3 19 4 AxonArm Ergo 12K501 Table of contents...

Страница 5: ...of the elbow joint The Automatic Forearm Balance provides for harmonious movement control free swing and reduction of energy consumption It stores the energy that is released when the arm is extended...

Страница 6: ...gs 3 2 Conditions of use The product is designed for use by unilateral or bilateral amputees The product is intended exclusively for use on adults The product was developed for everyday use and must n...

Страница 7: ...ed theft prevention systems at the entrance exit of stores metal detectors body scanners for persons e g in airports or other sources of strong magnetic and electric al interference e g high voltage l...

Страница 8: ...high voltage lines transmitters transformer stations computer tomographs magnetic resonance tomo graphs etc Allow the patient to rest during the adjustment of the electrodes otherwise muscle fatigue w...

Страница 9: ...hetic glove e g 8S500 R L AxonSkin Natural when 8E500 Michelangelo hand is used Additional MyoBock components such as suction socket electrode electrode cable pull in tube and tube valve for suction s...

Страница 10: ...long co contraction on channels 1 and 2 of the AxonMaster releases or locks the lock A short co contraction switches the gripping mode of the gripper A pulse on channel 3 of the AxonMaster releases or...

Страница 11: ...Reduction in mechanical strength due to thread misalignment The circuit board and the edge of the lamination ring must be parallel to each other Left picture The EasyPlug has been snapped in properly...

Страница 12: ...isted 7 Slide the strap clamp under the lamination ring 8 Screw the strap clamp in 6 5 Setting the Preflexion The level of AFB compensation flexion assistance can be adjusted and adapted to the weight...

Страница 13: ...e forearm or the upper arm socket due to released lock The Axon Bus prosthethic system must always be flexed for donning and doffing The Axon Bus prosthethic system must be stored in flexed condition...

Страница 14: ...rntable friction is easily adjusted by turning of the external adjustment screw using a coin or similar object 6 7 Installing the Pull Cable CAUTION Removing the pull cable Injury due to malfunction o...

Страница 15: ...ing Receptacle The charging receptacle Fig 1 item 2 has the following functions Contacts for charging the battery LED display to indicate the current battery charge level LED display for providing fee...

Страница 16: ...1x short 1x long Beep Pulsating Turn the Axon Bus pros thesis solution on Push the button for one to four seconds Turn the Axon Bus pros thesis solution off Push the button for more than one second C...

Страница 17: ...that may cause undesired operation This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to part 15 of the FCC Rules These limits are designed to pr...

Страница 18: ...ing General AxonArm Ergo Reference number 12K501 Dimensions mm Approx 345 mm x d55 mm to d80 mm arm extended Weight without Axon Bus terminal device g Approx 840 Forearm length as delivered mm 305 Min...

Страница 19: ...ds Probable cause The Axon Bus prosthetic system is overheated Remedy 1 Allow the Axon Bus prosthetic system to cool for a few minutes 2 Do not use the Axon Bus prosthetic system at high ambient tempe...

Страница 20: ...sion SB_2021 02 FM483 SB_210x297 Otto Bock Healthcare Products GmbH Brehmstra e 16 1110 Wien Austria T 43 1 523 37 86 F 43 1 523 22 64 info austria ottobock com www ottobock com Ottobock 647G790 0 en_...

Отзывы: