Ottobock | 17
Required accessories for operation
•
1 pc. battery charger 757L35 MyoCharge Integral
•
757VAK1 battery adapter cable (only for combination with VASI system)
6 Preparation for use
6.1 Required Tools
•
Screwdriver
•
Adhesive tape
•
726W9=* conical drill
•
Hand drill ø 4 mm
6.2 Recommended materials
•
623T1* nylon tube sock
•
99B71=* PVC sheeting
•
627B5=* Neoprene cell rubber tape
•
616T3=* ThermoLyn Trolene
•
200 617H12=* Pedilen rigid foam
•
617P21=* Pedilen hardener
•
99B81=* PVA bag
•
623T* Perlon stockinette
•
616G12 carbon fibre cloth
•
616G12 carbon fibre cloth
6.3 Laminating instructions
The product can be positioned in the upper arm or forearm socket. Depending on how much
space is available, the battery is secured in the outer socket or between the outer and inner sock
et.
NOTICE
Use of the product with a damaged flat cable
Malfunction due to incorrect or interrupted data transfer.
►
Avoid bending or pressing the flat cable multiple times in the same position.
INFORMATION
The length of the flexible flat cable between the battery and the charging receptacle can be
adjusted by carefully bending or folding it. Only light pressure must be applied when pinching
and sliding down the folded edge. If the flexible flat cable is bent in the opposite direction at a
spot where it has already been folded, serious damage can occur.
6.3.1 Laminating instructions for the upper arm socket
1) Position the battery as far proximal as possible on the alignment aid for the ErgoArm elbow
component.
2) Use the cap screws to screw the battery to the alignment aid from the inside (see fig. 2).
3) Connect the 2-pin supply cable to the electronics housing of the elbow joint.
INFORMATION: The 3-pin communication cable is not plugged in.
6.3.2 Laminating instructions for the forearm socket
1) Fabricate the inner socket using the familiar technique.
2) Cut nylon tube sock to twice the length, pull one half over the inner socket and tie it off.
3) Establish the position of the battery dummies (on the inner socket for patients with a long
residual limb).
4) Pull the rest of the nylon tube sock down so it is free of wrinkles and tie off distally.
5) Warm the PVC sheeting and pull it over the residual limb.
6) Tie off the PVC sheeting and evacuate.
7) Glue on neoprene cell rubber tape in 2 layers according to the course of the outer socket
edge.
Содержание 757B35-1
Страница 2: ...1 2 3 4 5 6 7 8 2 Ottobock...
Страница 3: ...9 10 11 12 13 14 15 16 Ottobock 3...
Страница 104: ...104 Ottobock MyoEnergy Integral 757B35 2 2 1 2 1 1 1 1 LED LED 1 2 1 3 1 4 3 3 1...
Страница 105: ...Ottobock 105 3 2 3 3 112 3 4 4 4 1 4 2 1 2 4 3...
Страница 106: ...106 Ottobock Ottobock DermaClean 453H10 1 4 4 112...
Страница 111: ...Ottobock 111 3 9 10 10 1 10 2 10 3 CE 93 42 VII 11 11 1...
Страница 114: ...114 Ottobock 2 2 1 2 1 1 1 1 1 2 2 1 3 1 4 3 3 1 3 2 3 3 122 3 4...
Страница 115: ...Ottobock 115 4 4 1 4 2 1 2 4 3 Ottobock DermaClean 453H10 1...
Страница 116: ...116 Ottobock 4 4 122...
Страница 121: ...Ottobock 121 2 3 9 10 10 1 10 2 10 3 93 42 IX I VII 11 11 1...