15
►
Follow the instructions regarding the length of the screws and about how to secure the
screws.
5.1 Adjusting the vacuum pump
Vacuum generation and damping of the vacuum pump can be adjusted using the elastomer rod.
Plantar flexion and dorsiflexion as well as pronation and supination can be adjusted via the pyram
id receiver.
When a socket adapter with rotatable pyramid receiver is installed, this can be used for rotation.
Adjusting the damping
>
Required materials:
Allen wrench 4 mm
1) Disassemble existing distal prosthetic components.
2) Screw in the adjustment screw in the distal end of the vacuum pump fully to the stop (see
fig. 3).
3) Loosen the adjustment screw as described in the adjustment table below.
4)
Optional:
Fully unscrew the adjustment screw to replace the elastomer rod (see fig. 4).
Unscrew the set screws of the pyramid receiver sufficiently in order to do so. After inserting a
new elastomer rod, complete the adjustment procedure again.
5) Install the distal prosthetic components as described in the respective instructions for use.
Adjustment table (turning out the elastomer rod)
Body weight [kg]
Product
Red rod
Yellow rod
50 (110 lbs)
3
–
60 to 70 (132 lbs to 154 lbs)
2.5
–
80 to 90 (176 lbs to 198 lbs)
2
–
100 (220 lbs)
4R180, 4R181, 4R182,
4R183
1.5
3
110 (242 lbs)
–
3
120 to 130 (264 lbs to 286 lbs)
–
2.5
140 to 150 (308 lbs to 330 lbs)
4R181, 4R183
–
2
The optimal setting depends on the user's level of activity and requests.
Loosen the adjustment screw of
the elastomer rod by a maximum of four (4) turns.
5.2 Installing the vacuum pump
Countersunk head screws that must be used for installing a socket adapter are included with the
vacuum pump. The countersunk head screws are chosen according to the socket adapter.
Socket adapter
Compatible countersunk head screw
Pyramid/pyramid receiver, fixed
501S41=M6x12-12.9
Pyramid/pyramid receiver, rotatable
501S41=M6x16
>
Required materials:
torque wrench (e.g. 710D20), 636K13 Loctite 241
1) Select the countersunk head screws that are compatible with the socket adapter (see selec
tion table).
2)
INFORMATION: If the pyramid receiver or the pyramid is rotatable: Place the pressure
plate on the adapter.
Position the socket adapter.
3) Secure the screws with Loctite.
4) Tighten the posterior screws first (
12 Nm
), then tighten the anterior screws (
12 Nm
).
5) Install the other prosthetic components as described in the respective instructions for use.
6) Carry out the static alignment of the prosthesis.
INFORMATION: The length of the vacuum hose can be determined once all prosthetic
components are correctly aligned.
Содержание 4R182 Harmony P4 modular
Страница 2: ...1 Loctite 241 12 Nm 501S41 M6x12 12 9 501S41 M6x16 2 2...
Страница 3: ...3 4 5 6 7 3...
Страница 110: ...110 2 3 20 C 60 C 20 90 10 C 45 C 20 90 2 4 ISO 10328 3 3 5 2 5 Harmony 2 6 3 3 1 3 2 109...
Страница 111: ...111 4 1 1 1 1 Harmony P4 modular 2 1 4X632 3 4 501S41 M6x12 12 9 4 4 501S41 M6x16...
Страница 114: ...114 5 4 2 3 1 1 2 3 1 2 3 3 2 4 500 850 2 2 4 500 850 2 500 850...
Страница 115: ...115 500 850 6 6 1 1 2 451F21 3 6 2 1 2 7 7 1 7 2 7 3 5 1 2 3 4 1...
Страница 116: ...116 2 3 4 60 5 6 7 8 8 30 112 9 10 10 1 10 2 2017 745 CE 10 3...
Страница 118: ...118 2 3 20 C 60 C 20 90 10 C 45 C 20 90 2 4 ISO 10328 300 3 5 2 5 Harmony TBS TBS 2 6 3 3 1 3 2 117...
Страница 123: ...123 5 1 2 3 4 1 2 3 4 60 ml 5 6 7 8 8 30 1 120 9 10 10 1 10 2 EU 2017 745 CE 10 3...
Страница 124: ...124 11 4R182 4R183 g 545 665 mm 125 kg 50 100 110 220 50 150 110 330 2 4...
Страница 125: ...125...
Страница 126: ...126...
Страница 127: ...127...