If assistance is required during the start-up, use or maintenance of the product, as well as in the event of
unexpected operating behaviour or circumstances, contact the Ottobock office responsible for your country.
Type of
error
Event
User restriction Required action
Warning
e.g. service interval has
expired
No restriction
Consult Ottobock
Hydraulic
system
overheated
Overheating of the
hydraulic unit e.g. due to
uninterrupted, increased
activity
Reduced activity
Allow the hydraulic unit
to cool down by reducing
activity
Error
e.g. a sensor
(tube adapter, triple sen-
sor, ...) is not operating
Walking possible with
restrictions
Restart the knee joint by
attaching and removing the
inductive charger (up to
three times);
If error remains, consult
Ottobock immediately
Selftest failure
(if signals occur after
removing the charger)
AXON
Tube Adapter
disconnected
AXON Tube Adapter has
been disconnected while
the knee joint is activated
and the inductive
charger is not attached
Connect the
AXON Tube Adapter
AXON Tube Adapter
not connected while
removing the inductive
charger
Critical error
e.g. failure of the valve
drive
High stance resistance
Swing phase no longer
possible
Restart the knee joint by
attaching and removing the
inductive charger (up to
three times);
If error remains, consult
Ottobock immediately
Selftest failure
(if signals occur after
removing the charger)
Charging
Status 0%
No electronic
functionality
Activation of safety mode
Swing phase no longer
possible
Charging required
No functionality
Consult Ottobock
immediately
Error code on
the display of
the Remote
Control
Consult Ottobock
immediately
(Error code number
required)