Check the product settings using the corresponding adjustment software of the
Ottobock Data Station.
17.2 Operating states/error signals
The prosthesis indicates operating states and error messages through beeps and vibration signals.
17.2.1 Signals for operating states
Battery charger connected/disconnected
Beep signal
Vibration signal
Event
3x long
Charging mode started (3 sec. after connecting the bat
tery charger)
1x short
1x before beep signal
Self-test completed successfully, product is operational
Mode switching
INFORMATION
When mute mode (silent mode) is activated, no beep and vibration signals are generated.
INFORMATION
If the
Volume
parameter is set to '0' in the Cockpit app, there are no beep signals (see Page 93).
Beep signal
Vibration sig
nal
Additional action performed
Event
1x short
1x short
Mode switching using the Cockpit
app
Mode switching is performed using the
Cockpit app.
1x long
1x long
Bouncing on the forefoot followed
by unloading the prosthetic leg
Bouncing pattern recognised.
1x short
1x short
Weight taken off prosthetic leg
and leg kept still for 1 second
Switching to basic mode (mode 1) car
ried out.
2x short
2x short
Weight taken off prosthetic leg
and leg kept still for 1 second
Switching to MyMode 1 (mode 2) car
ried out.
3x short
3x short
Weight taken off prosthetic leg
and leg kept still for 1 second
Switching to MyMode 2 (mode 3) car
ried out.
17.2.2 Warnings/error signals
Error during use
Beep signal
Vibration signal
Result
Required action
–
1 x long at intervals of
approx.
5 seconds
(if
mute mode (silent mode)
is activated, this signal is
not generated)
Overheated hydraulic unit Reduce activity.
–
3 x long
Charge level under 25%
Charge battery soon.
Remaining
operating
time
approx. 24 hours
–
5 x long
Charge level under 10%
Charge battery soon
Remaining
operating
time
approx. 6 hours
5 x long
5 x long repeated every
60 seconds
Medium
error
(see
Page 100)
e.g. a sensor is not oper
ational
Walking possible with restric
tions. Please note the change
in flexion resistance.
The product must be inspec
ted immediately by an author
ised Ottobock Service Center.
107
Appendices
Содержание 3B1-2/3B1-2-ST
Страница 4: ...min 3 mm min 5 mm 7 1 8 5 sec 1 2 3 7 1 1 3 Connection Job Specification Alignment Basic Mode 7 1 6 4 ...
Страница 62: ...62 ...
Страница 115: ...115 ...