42
16 Erasing the data memory
Please note:
After erasing the data memory, the measured values stored in the
OTT ecoLog 500 (3G) are permanently lost! If necessary download measured
values before erasing!
How to erase the data memory
Start the operating program.
Establish the PC/OTT ecoLog 500 (3G) communication link, refer to Chapters
9.2 or 9.3.
Click the "Setup device" button in the start window.
Select the "Delete data memory" function in the "OTT ecoLog 500 (3G)" menu.
Acknowledge the "Are you sure you want to delete data memory?" warning by
clicking "Yes"
➝
The operating program erases the complete data memory of
the OTT ecoLog 500 (3G) (all measurement channels including the info chan-
nel). The OTT ecoLog 500 (3G) then determines and stores the water level and
the water temperature again using the sample interval settings set.
17 Updating the OTT ecoLog 500 (3G) firmware
If required, you have the option of updating the OTT ecoLog 500 (3G) firmware
(operating system software). This makes sense, if e.g. devices delivered at different
times are to receive the same operating system version. Updating is carried out
via the operating program. According to availability, an updated version of the
OTT ecoLog 500 (3G) firmware can be found on the internet site "www.ott.com"
in the "myOTT" area.
How to update the firmware
Download the new version of the firmware (File: e.g. "5544630293_01035_ -
el500NG.bin" or 5544830293_01000_el500NG.bin) from the internet site.
Move the file "5544630293_XXXXX_el500NG.bin" or "5544830293_XXXXX -
_el500NG.bin" to the directory, in which the operating program is located.
Establish the PC/OTT ecoLog 500 (3G) communication link, refer to Chapters
9.2 or 9.3.
From the "OTT ecoLog 500 (3G)" menu, select the "Firmware update" function.
Confirm message by clicking "Yes"
➝
The operating program copies the new
firmware to the OTT ecoLog 500 (3G). The OTT ecoLog 500 (3G) then deter-
mines and stores the water level and the water temperature again using the
sample interval settings set.
Please note:
During updating process, avoid breaking the communication
link (e.g. by accidentally removing the OTT DuoLink from the infrared interface
of the OTT ecoLog 500 (3G)). If the communication link is aborted, the
firmware will no longer be executable! Similarly, no other programs should be
started or files opened during the copying process!
Notes
If there are multiple ".bin" files in the directory you will have to select the
required file manually.
The measurements saved in the OTT ecoLog 500 (3G) are not lost after an
update.