Short increment indication
•
The presence of a short increment in intensity is indicated by the little
L
or
R
test ear indicator in the display changing
briefly to a rectangle,
.
•
To change the number of small increments turn the
Right level
knob.
Manual scoring of responses
•
If the patient is unable to use the patient responder, but registers an increment heard in another way, you can press
the
Right Present Signal
button to indicate that the tone was heard.
Pausing the test
•
You can pause the SISI test by pressing
Start/Pause
again. During the pause "SISI" will flash in the display. To resume
the test press
Start/Pause
once more.
Stopping the test
•
During a pause, you can
stop the test completely
by pressing
More
.
Changing the dB level
•
To change the dB level use the
left LEVEL
knob.
Changing the dB increments
•
To change the dB increments use the
Right level
knob (0.25 - 5dB).
If you do so while the test is in progress, the counters are reset to zero and the test starts again from the beginning.
Changing number of increments
•
To change the preset number of SISI dB increments transmitted during a test (10-50) press
Setup
once and use the
Right level
knob. Then press
Setup
again to implement the new setting. While the SISI test is in progress, it is not pos-
sible to change the frequency or the number of dB increments.
7.6.1
Masking during SISI
Masking
To perform masking, press
Contra Air
.
•
When you hold down the key, the masking intensity (instead of % score) and signal type are displayed.
Intensity
•
To change the intensity press and hold the key and turn the
Right level
knob
Signal type
•
To change the signal type press and hold the key and turn the
left LEVEL
knob.
7.6.2
Using STORE in the SISI test
Note
•
A maximum of four different sets of SISI values can be stored for both left and right ear.
Otometrics - MADSEN Xeta
75
7 Testing a patient with MADSEN Xeta
Содержание Madsen Xeta
Страница 1: ...MADSEN Xeta Reference Manual Doc No 7 50 0870 EN 17 Part No 7 50 08700 EN ...
Страница 6: ...Index 103 6 Otometrics MADSEN Xeta ...
Страница 26: ...3 Overview of MADSEN Xeta 26 Otometrics MADSEN Xeta ...
Страница 44: ...4 Navigating in the OTOsuite Audiometry Module 44 Otometrics MADSEN Xeta ...
Страница 54: ...5 Preparing for testing 54 Otometrics MADSEN Xeta ...
Страница 66: ...6 MADSEN Xeta Test Settings 66 Otometrics MADSEN Xeta ...
Страница 86: ...11 Unpacking and installing 86 Otometrics MADSEN Xeta ...
Страница 88: ...12 Maximum non destructive voltage 88 Otometrics MADSEN Xeta ...
Страница 94: ...14 Standards and safety 94 Otometrics MADSEN Xeta ...
Страница 102: ...15 Technical specifications 102 Otometrics MADSEN Xeta ...