Level knob
Level knob
Controls the functions of the
Left
and
Right level
knobs:
Always use left
The
Left level
knob controls the signal and the
Right level
knob controls mask-
ing.
Always use right
The
Right level
knob controls the signal and the
Left level
knob controls mask-
ing.
Follow display
The stimuli (signal/masking) controlled by the
Left
and
Right level
knobs cor-
respond to the placing (left/right) of the stimuli on the display.
Auto delete
Auto delete
Automatically deletes the oldest patient record to free memory space as
required.
Default
On
AC transducer
AC Transducer
Choose between calibrated headphones.
Note
•
If the wrong headphone is selected compared to the one that is phys-
ically con-nected to Xeta, the measurements will be incorrect!
•
- - - (None)
•
TDH39 (Use this setting also for ME 70 transducers)
•
EAR3A (E-A-R-TONE® 3A)
BC transducer
BC transducer
Choose between calibrated bone conductors.
•
- - - (None)
•
B-71M (Mastoid)
Pulse frequency
Pulse frequency
Sets the pulse frequency:
•
0.25 to 2.50 Hz
62
Otometrics - MADSEN Xeta
6 MADSEN Xeta Test Settings
Содержание Madsen Xeta
Страница 1: ...MADSEN Xeta Reference Manual Doc No 7 50 0870 EN 17 Part No 7 50 08700 EN ...
Страница 6: ...Index 103 6 Otometrics MADSEN Xeta ...
Страница 26: ...3 Overview of MADSEN Xeta 26 Otometrics MADSEN Xeta ...
Страница 44: ...4 Navigating in the OTOsuite Audiometry Module 44 Otometrics MADSEN Xeta ...
Страница 54: ...5 Preparing for testing 54 Otometrics MADSEN Xeta ...
Страница 66: ...6 MADSEN Xeta Test Settings 66 Otometrics MADSEN Xeta ...
Страница 86: ...11 Unpacking and installing 86 Otometrics MADSEN Xeta ...
Страница 88: ...12 Maximum non destructive voltage 88 Otometrics MADSEN Xeta ...
Страница 94: ...14 Standards and safety 94 Otometrics MADSEN Xeta ...
Страница 102: ...15 Technical specifications 102 Otometrics MADSEN Xeta ...