MONO INSERT
Connection for the black 1/4" jack plug from the EAR-3A mono insert phone.
Patient Response switch
Connection for the black 1/4" jack plug from the patient responder.
RS 232 interface
Connection for Xeta to a PC. 9-pole PC- RS232 interface cable.
The installation must be carried out in accordance with Medical Electrical Systems in IEC
60601-1 3.1 edition: 2012, ANSI/AAMI ES60601-1 (2005) + AMD 1 (2012) and CAN/CSA-
C22.2 No. 60601-1 (2014). The supplementary provisions on the reliability of electro-med-
ical systems.
It is a general rule for all electrical equipment used in the proximity of the patient that:
•
The connected equipment must comply with Medical Electrical Systems in IEC 60601-1 3.1:edition:2012.
Any PC connected to the Xeta must comply with the requirements of IEC 60950, "Safety of information technology equip-
ment, including electrical business equipment".
Supply voltage connection
100-120 V AC, 200-240 V AC, 50/60 Hz, 60 VA
Fuses: 2 T 1 A H/250 V, 5 mm x 20 mm.
Warning
•
Before you replace a fuse, see
Power ON/OFF
On/Off switch.
Otometrics - MADSEN Xeta
25
3 Overview of MADSEN Xeta
Содержание Madsen Xeta
Страница 1: ...MADSEN Xeta Reference Manual Doc No 7 50 0870 EN 17 Part No 7 50 08700 EN ...
Страница 6: ...Index 103 6 Otometrics MADSEN Xeta ...
Страница 26: ...3 Overview of MADSEN Xeta 26 Otometrics MADSEN Xeta ...
Страница 44: ...4 Navigating in the OTOsuite Audiometry Module 44 Otometrics MADSEN Xeta ...
Страница 54: ...5 Preparing for testing 54 Otometrics MADSEN Xeta ...
Страница 66: ...6 MADSEN Xeta Test Settings 66 Otometrics MADSEN Xeta ...
Страница 86: ...11 Unpacking and installing 86 Otometrics MADSEN Xeta ...
Страница 88: ...12 Maximum non destructive voltage 88 Otometrics MADSEN Xeta ...
Страница 94: ...14 Standards and safety 94 Otometrics MADSEN Xeta ...
Страница 102: ...15 Technical specifications 102 Otometrics MADSEN Xeta ...