19
Technical specifications
Type identification
MADSEN OTOflex 100 is type 1012 from GN Otometrics A/S
Compliance measuring system
Probe tone:
226 Hz at 85dB SPL ± 1.5 dB
1000Hz at 75dB SPL ± 1.5 dB
THD:
< 3% in 2 cc
Frequency accuracy:
±0.5%
Range:
0.1 ml to 8.0 ml ± 5% or 0.1 ml, whichever is greater
Acoustic reflex
Contralateral Stimulation
Pure tones:
500 Hz, 1000 Hz, 2000 Hz, 3000 Hz, 4000 Hz
Frequency accuracy:
± 0.5%
Noise
White Noise according to IEC 1027
Low Pass 400 to 1600 Hz.
High Pass 1600 to 4000 Hz.
Roll off >12 dB/Octave.
Range at:
BBN, LPN at 50 to 100 dB HL ±3 dB
HPN at 50 to 95 dB HL ±3 dB
Step size dB
1, 2, 5, 10 dB
E-A-RTONE® 3A:
Range at:
500 Hz at 50 to 105 dB HL ± 3 dB
1000 Hz at 50 to 120 dB HL ± 3 dB
2000 Hz at 50 to 115 dB HL ± 3 dB
3000 Hz at 50 to 105 dB HL ± 3 dB
4000 Hz at 50 to 110 dB HL ± 3 dB
THD:
< 3% in 2 cc (measured 5 dB below max output)
Immittance probe:
Range at:
500 Hz at 50 to 105 dB HL ± 3 dB
1000 Hz at 50 to 120 dB HL ± 3 dB
2000 Hz at 50 to 115 dB HL ± 3 dB
3000 Hz at 50 to 105 dB HL ± 3 dB
4000 Hz at 50 to 110 dB HL ± 3dB
THD:
< 3% in 2 cc (measured 5 dB below max output)
176
Otometrics - MADSEN OTOflex 100
19 Technical specifications