9. Continue to use the right and/or left
LEVEL
knob to increase/decrease intensities until you have determined all
thresholds.
Note
•
If the thresholds measured at 1000 Hz in the ear differ by 10 dB or more, even the other frequencies should
be retested.
10. Continue to press
STORE
after each threshold is found to store that particular threshold.
11. After each threshold is determined and stored, you will notice that as you turn the
LEVEL
knob to scroll frequencies,
each stored threshold at that particular frequency will be marked by an asterisk (*) and the relevant audiogram symbol
indicating the stored threshold.
12. If you are not performing the test online with the OTOsuite Audiometry Module, plot the results in an audiogram
sheet.
13. Repeat this procedure for the other ear.
Note
•
If there are differences between the thresholds in the two ears, see
.
6.1.2
Sound Field testing
When you select SF transducer, and Free Field speakers (F.F.) are assigned as transducers (
SETUP
,
), the
results are stored as an individual threshold curve using the applicable Sound Field symbols (
).
The symbol used for Sound Field applies to both ears, unless masking has been used.
Tip:
If ear-specific data is retrieved by blocking the contra-lateral ear, activate the masking on the audiometer to use the cor-
rect symbols. This is especially important when you use MADSEN Itera II with the OTOsuite Audiometry Module.
Set the masking level to e.g. -10 dB, and note in the Comments field that the contra-lateral ear was blocked.
6.2
Bone Conduction threshold test
Note
•
In Bone Conduction testing, masking is always necessary in order to determine in which ear the presented tone
signal is actually heard. This is because of the reduced level of inter-aural attenuation.
Note
•
Make sure that there are no interfering ambient sounds, particularly at low frequencies. If necessary, place the
patient in a sound booth.
Prerequisites
•
Press
TONE
.
•
Press
BC
.
•
If needed, select
WARBLE
or
PULSE
as stimulus type.
62
Otometrics - MADSEN Itera II
6 Tone testing
Содержание madsen itera ii
Страница 1: ...MADSEN Itera II Reference Manual Doc No 7 50 0860 EN 26 Part No 7 50 08600 EN ...
Страница 10: ...1 Introduction to MADSEN Itera II 10 Otometrics MADSEN Itera II ...
Страница 52: ...4 Navigating in the OTOsuiteAudiometry Module 52 Otometrics MADSEN Itera II ...
Страница 60: ...5 Preparing for testing 60 Otometrics MADSEN Itera II ...
Страница 68: ...6 Tone testing 68 Otometrics MADSEN Itera II ...
Страница 74: ...7 Speech testing 74 Otometrics MADSEN Itera II ...
Страница 86: ...9 Managing Data and Results 86 Otometrics MADSEN Itera II ...
Страница 110: ...15 Maximum non destructive voltage 110 Otometrics MADSEN Itera II ...
Страница 118: ...18 Standards and safety 118 Otometrics MADSEN Itera II ...