Maintenance and calibration
MADSEN Itera II requires no preventive maintenance except for regular calibration of the transducers.
See
.
13.1
Service and repair
It is recommended that you keep the packing material in which MADSEN Itera II was delivered. If you need to send it in
for service, the original packing material will ensure protection against damage during transport, etc.
Warning
•
For the sake of safety and in order not to void the warranty, service and repair of electro-medical equip-
ment should be carried out only by the equipment manufacturer or by service personnel at authorized workshops. In
case of any defects, make a detailed description of the defect(s) and contact your supplier. Do not use a defective
device.
Note
•
There are no user-serviceable parts inside the MADSEN Itera II cabinet.
On request, your supplier can obtain a service manual from the factory. The service manual contains electrical diagrams,
descriptions, lists of components and calibration instructions, etc.
13.1.1
Fuses
To replace or inspect a fuse, proceed as explained below.
Warning
•
Risk of fire
. Before replacing a fuse, first switch off the instrument and disconnect it from the
mains power supply.
The fuse-holders are located next to the AC power connector (or they are integrated with the power connector).
To remove a fuse:
•
Unscrew the end of the fuse-holder using a screwdriver with a blade at least 6 mm wide (or remove the power con-
nector).
Fuse types
The fuse is a push-fit in the end of the fuse-holder. When you replace fuses, use only fuses of type T 1 A H/250 V, 5 mm x
20 mm.
13.2
Cleaning
Cleaning the device
There are no specific requirements to sterilization or disinfection of the device.
Make sure that the instrument is kept clean and free of dust:
•
Remove dust using a soft brush.
•
To clean the cabinet, use a soft, slightly damp cloth with a small amount of mild detergent on it.
Otometrics - MADSEN Itera II
103
Содержание madsen itera ii
Страница 1: ...MADSEN Itera II Reference Manual Doc No 7 50 0860 EN 26 Part No 7 50 08600 EN ...
Страница 10: ...1 Introduction to MADSEN Itera II 10 Otometrics MADSEN Itera II ...
Страница 52: ...4 Navigating in the OTOsuiteAudiometry Module 52 Otometrics MADSEN Itera II ...
Страница 60: ...5 Preparing for testing 60 Otometrics MADSEN Itera II ...
Страница 68: ...6 Tone testing 68 Otometrics MADSEN Itera II ...
Страница 74: ...7 Speech testing 74 Otometrics MADSEN Itera II ...
Страница 86: ...9 Managing Data and Results 86 Otometrics MADSEN Itera II ...
Страница 110: ...15 Maximum non destructive voltage 110 Otometrics MADSEN Itera II ...
Страница 118: ...18 Standards and safety 118 Otometrics MADSEN Itera II ...