Quick Guide:
Bio-logic
®
NavPRO ONE
®
ABR screening
Otometrics - Next generation Bio-logic products
4
Bio-logic
®
Select the appropriate ear tip size and place it on the probe tip. Insert the probe in the ear making
sure that the ear tip fits tightly into the ear canal.
Select QuickABR button
Select the test ear by pressing
either the Right or Left button
A calibration will be performed
prior to measurement.
A green light indicates
successful calibration.
An impedance check
will be performed prior to
measurement. An impedance
below 6 k ohm is recommended
for best test results.
After successful ear probe
calibration and impedance
check, the test will start
automatically. If yellow or
red lights are displayed, check
to make sure you are using the
correct probe tip and ear tip
size, that the eartip is inserted
correctly, and that there is no
excessive ambient noise.