Cleaning AccuScreen
•
Before cleaning, switch off AccuScreen and unplug it from any external power source.
•
Unplug the cables from AccuScreen.
•
Use a tissue moistened with cleaning fluid to clean the surface of the device.
To maximise the housing lifetime, avoid the following chemicals:
•
Isopropyl, 70%
•
Ethanol, >40% concentration
•
Formaldehyde
•
Dichloro-meta-xylenol 5-10%
If disinfecting AccuScreen is considered necessary, some suitable alternatives are cleaning agents that contain:
•
Hydrogen peroxide, 3%, 30%
•
Peracetic acid 0.5-5%
•
Sodium hypochlorite 1-10%
•
Glutaraldehyde 2-5%
•
Ortho-phthalaldehyde (a.k.a. OPA) 0.5-2%
•
Chlorhexidingluconat 2-4%
Caution
•
Always make sure that no moisture enters the probe, the sockets (data interface connector and probe plugs)
or the test cavity.
Caution
•
Never immerse AccuScreen into water or other cleaning solutions.
Caution
•
Use of cleaning agents other than those recommended in the user manual can cause damage to the device
(for example, stress cracks in molded plastic).
11.4.1
Cleaning and disinfecting the built-in test cavity
If needed, use a tissue moistened with any of the recommended disinfectant agents described in section 11.4 to clean the
entry surface of the test cavity.
Caution
•
Make sure that no liquid enters the test cavity.
If the test cavity has been contaminated with debris from the probe tip, make sure that the test cavity cannot be used, for
instance by sticking a piece of tape across the entry hole, and contact your authorised service department for cleaning
and/or replacement of the test cavity.