217
20.3
ABR cable test
The ABR cable test checks the continuity of the electrode cables.
Attach the probe snaps to the top of the cable tester as illustrated. Select
ABR Cable Test
and wait for the result to be displayed. If the electrode
cables are OK then a large tick is displayed on the screen. If any of the
cables fails the test, a large cross is displayed.
Select
Details
to see the results for each cable. The impedances should
show zero for all cables.
If any cable shows a cross, check that it is properly connected to the
Otoport and retest. If it continues to fail, then replace the cable.
QA
area
Otoport OAE+ABR
User Manual for Advance and Screener