ABR test procedure
7
ABR test
Use the left or right arrow key to choose your ABR test screen.
Note:
The ABR test can operate in different Modes. The following text
refers to Screening Mode, where Autostart is ON. See section
10.3.4
in the
Configuration
chapter for Mode configuration
option. The electrode colours are for the standard electrode
connection cable and may differ if other cables are in use.
Connect the electrode leads and the probe to the ABR unit and prepare
your patient. Connect the red (+ve) cable to the high forehead, the white
(-ve) cable to the nape of the neck and the black (common) cable to the
back of the shoulder (see chapter
5
Test preparation
for full details).
Select
Test
to being testing.
Note:
If you see the message ABR Mains is Not Set”, your Otocheck
has not yet been permanently configured for your region. You
must select the powerline frequency in your region (see
ABR
Mains
in
10.3
) Test Settings).