Страница 1: ...ID 1588 ver 205 1...

Страница 2: ...EMORY LOCATION 7 4 2 CLEARING THE ENTIRE MEMORY 7 5 USING THE PRINT OR TRANSMIT FUNCTION 7 6 DISABLING REFLEX FREQUENCIES 10 7 CALIBRATING THE PROBE 11 8 CHOOSING NUMBER OF TYMPANOGRAM SWEEPS 12 9 HOW...

Страница 3: ...t using an external printer connected to the tympanometer or transmitted to a PC through a serial cable TSM 300 may store results from 20 tests For configuring and calibrating the device there is a me...

Страница 4: ...ating the probe 2 2 Taking a measure Before taking a test on a patient mount a new rubber ear tip on the probe that fits the patient s ear size Ensure the patient s ear canal is clean and not clogged...

Страница 5: ...Hz are presented as short bursts If a change in compliance greater than 0 05 ml is detected a reflex is considered present Because this is an extremely small compliance change any movement of the prob...

Страница 6: ...l the menu Left button Used to go back in the menu or adjust a setting Down button Moves the menu selection bar down Up button Moves the menu selection bar up Right button Used to enter the selected m...

Страница 7: ...red in the memory can be exported in two ways They may be printed out on a printer connected directly to the tympanometer or they may be transmitted to a connected PC This is done by pressing the righ...

Страница 8: ...he printer is a Kyosha thermo printer This setting may work with other small printers too Select HP if the printer supports PCL 2 to PCL 5 language Some printers may emulate other printers For example...

Страница 9: ...ut The Test printer item will print out a test image with the selected margin To leave the menu press MENU Clear data Clear all data Printer setup Printer type Reflex setup Printer Port Calib cavity C...

Страница 10: ...happen if an IBM Pro Printer is selected in the setup and an Epson compatible printer is used or vice versa If the printer just advances the paper without printing anything it may help to enable Carri...

Страница 11: ...p into the hole of the test cavity labelled 2 0 ml and then press Right key 5 Press the Down key to select 5 0 ml 6 Put the probe tip with ear tip into the hole of the test cavity labelled 5 0 ml and...

Страница 12: ...s Clear data Clear all data Printer setup Reflex setup Calib cavity Right key Tymp sweeps 1 If you want a quick test without repeatedly sweeping and graph redrawing set the value to 1 The device will...

Страница 13: ...there is ear wax or other things in the probe head then the probe can be dismounted for better cleaning Please do as shown in the next figures Probes handle Probe head To remove the probe head hold f...

Страница 14: ...Another difference is that the 226Hz measuring tone is kept at a constant signal level instead of changing the level during a measure Often a feedback system is used to adjust the sound level dynamic...

Страница 15: ...flex measurements Test frequencies 500 Hz 1 kHz 2 kHz 3 kHz 4 kHz 2 Test method ipsilateral Test Level 105dB SPL in 2ml 3dB Supply 230 VAC 12 VDC 10 Dimension 225 mm x 180 mm x 55 mm Weight Approx 100...

Страница 16: ...ID 1588 ver 205 16 Warning and safety notices Maintenance calibration Cleaning Shipping recommendations Disposal...

Страница 17: ...ID 1588 ver 205 17 Symbols...

Страница 18: ...ID 1588 ver 205 18 EMC...

Страница 19: ...ID 1588 ver 205 19...