background image

O P T I M A

 

405-1307-001 Rev D

OrthoAccel®, AcceleDent®, SoftPulse Technology® are registered trademarks and 

Optima™ is a trademark of OrthoAccel Technologies, Inc.

For U.S. patent coverage, see: www.acceledent.com/patents 

DIRECTIONS FOR USE

93% RH

30% RH

1060 hPa

700 hPa   

50

Содержание AcceleDent OPTIMA

Страница 1: ...hoAccel AcceleDent SoftPulse Technology are registered trademarks and Optima is a trademark of OrthoAccel Technologies Inc For U S patent coverage see www acceledent com patents DIRECTIONS FOR USE 93...

Страница 2: ...celeDent Optima Device 5 3 Use schedule 5 4 AcceleDent App Use 6 0 Care 6 1 Cleaning 6 2 Storage 6 3 Disposal 7 0 Warranty 7 1 Express Limited Warranty 7 2 Legal Limitations 7 3 Customer Service and R...

Страница 3: ...ropulses to gently accelerate the movement of your teeth as they are guided by your orthodontics 2 0 About the Manufacturer Contact the manufacturer OrthoAccel about any questions or concerns that are...

Страница 4: ...s to help facilitate minor anterior tooth movement Contraindications for Use Use of osteoporosis drugs Poor oral hygiene Periodontal disease that is not under full control for at least 3 4 months prio...

Страница 5: ...symbol Check to make sure the AcceleDent Optima Activator is secure in the Charging Case for proper charging To close the lid securely push down on the front of the latch Lock symbol Note The lid does...

Страница 6: ...7 To pause your treatment session for up to 5 minutes press the on off button once During the pause the Status Indicator will pulse blue slowly When you are ready to resume press the on off button aga...

Страница 7: ...likely to forget a daily session 5 3 2 1 Some patients prefer using AcceleDent Optima just after they wake up and brush their teeth in the morning Others use the appliance while engaged in a consiste...

Страница 8: ...5 4 3 2 Follow specific steps recommended by your device manufacturer for downloading the app 5 4 4 Complete Registration 5 4 4 1 Open the app and begin registration by tapping on Register 5 4 4 2 En...

Страница 9: ...time You will then receive a push notification every day at this time even if your app is closed 5 4 10 Connect to Your Community 5 4 10 1 Within Community you can receive messages and keep track of f...

Страница 10: ...6 1 2 2 As needed unplug from power source and wipe external and internal surfaces using a soft moist cloth Dry with a soft cloth before reconnecting to power source 6 1 3 For the Mouthpiece 6 1 3 1 A...

Страница 11: ...nated collection facility for batteries and other electrical devices like cell phones computers video games and electronic tools 6 3 3 These designated collection facilities are free of charge to the...

Страница 12: ...ervice will require an RMA Return Merchandise Authorization To receive an RMA for obtaining warranty services you must call 1 866 866 4919 in North America or 44 0 203 318 1915 if outside of North Ame...

Страница 13: ...materials please contact the OrthoAccel Customer Service North America 1 866 866 4919 Outside North America 44 0 203 318 1915 Email customerservice orthoaccel com Online www acceledent com EMERGO EUR...

Страница 14: ...cause electric shock and permanent damage to your device This product is designed to operate within a range of 100 240 volts Voltage converters DO NOT guarantee voltage compatibility 8 5 2 WARNING NEV...

Страница 15: ...to operate this equipment 8 5 11 1 No modification of the equipment is allowed 8 5 11 2 DO NOT attempt to replace the battery 8 6 Precautions 8 6 1 DO NOT use AcceleDent Optima without it being presc...

Страница 16: ...questions concerns or to report an unexpected operation or events regarding your AcceleDent Optima product please contact the Customer Care Center for your country visit www acceledent com for contac...

Страница 17: ...combinations Medical Electrical Systems shall comply with the safety requirements stated in the general standard IEC 60601 1 edition 3 clause 16 Any equipment not complying with the leakage current re...

Страница 18: ...m immersion between 15 centimeters and 1 meter in depth 10 11 FCC Compliance Statement 10 11 1 This device complies with part 15 of the FCC Rules Operation is subject to the following two conditions 1...

Страница 19: ...e to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separation dis...

Страница 20: ...n such an environment 10 14 2 In accordance with Clause 4 3 of IEC 60601 1 AcceleDent Optima does not have Essential Performance Emissions Test Compliance The AcceleDent Optima uses RF energy for inte...

Страница 21: ...short interruptions and voltage variations on power supply input lines IEC 61000 4 11 100 drop 0 5 periods 0 45 90 135 180 225 270 315 100 dip 1 period 30 dip 25 30 periods Voltage Interruptions all...

Страница 22: ...quency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths f...

Страница 23: ...device to sale by or on the order of a physician Certification that the electromagnetic interference from the device is under limits approved by the Federal Communications Commission Class II Equipmen...

Страница 24: ...latch Close Charging Case by pushing down on this part of the latch Humidity limitation Atmospheric pressure limitation Not waterproof IP32 Is protected from intrusion of solid foreign objects greate...

Страница 25: ...France 33805080086 Italy 39800596833 Germany 498007244532 Outside North America 44 0 203 318 1915 www acceledent com Emergo Europe Prinsessegracht 20 2514 AP The Hague The Netherlands Phone 31 70 345...

Отзывы: