TECHNICAL NOTE 

 
 

Installing Sheaths on Tubing 19 and 20 

 
 
20/10/2005 

 

Purpose: To ensure a good fitting for tubing 19 and 20 on the counting heads. 
 
Concern: all mythic 18 SN :103804-000225  and lower 
 
Material: silicon Sheaths 
 
Labour time: 15 min 
 
 

1-

 

Drain the system [SERVICE] [HYDRAULICS] [DRAIN ALL], remove tubing 19 and 20, 
and disconnect the counting heads from the chambers. 

2-

 

With a tissue, wipe up the metallic fitting of the counting heads. 

3-

 

Prepare tubing 19 and 20 as follow: 

 

- Place the silicon sheath on one end of tubing 19 and 20. 
- Be sure that you have 3-4mm clearance of tubing 19 and 20. (Fig 1) 

  
                                             3-4 mm                                
 

                                        

   

 

Fig 1: tubing 19 and 20 not installed on the counting heads 

 

4- Connect tubing 19 on the metallic fitting of RBC counting head (side of the silicon sheath).  
5- Connect tubing 20 on the metallic fitting of WBC counting head (side of the silicon sheath), 
the other side on the RBC Chamber. (Fig2) 
 
 
 

                                                              
 
 
 
 
 

 

 

 

 

 

 

 

 

 

 

…/… 