Removing the liquid
Slide the small pipe, present on the bottom of the machine, out. Remove the cap (see photo); position the recipient under the manifold and wait for
the tank to empty completely. The circuit contains approx. 2 litres of liquid.
Topping-up the liquid
Re-position the previously-removed cap on the bottom of the manifold, using the wrench supplied. Therefore, introduce the syringe with
double-distilled water into one of the pipes until the tank is filled completely.
WARNING!
Before closing the compartment, check the presence of the caps on the 2 small pipes
Checking the filters
There are two air filters on the bottom of the machine, which must be cleaned every 6
months (Ref 1). To do this, loosen the grid clamping screws and clean using compressed
air. If the filter is too dirty or damaged, replace it (Code RIC00040).
Periodically check the active carbon filter and replace it at least once every 6 months (Ref
2), loosen the two fixing pawls and remove downwards, puling by the wings.
WARNING!
Do not cover the equipment vents with cloths or other materials. Leave a suitable space of at least 10 cm between the rear vent and
the wall. If this is not the case, equipment efficiency will be compromised.
Replacing the air filter
Loosen the 2
screws shown in the photo, slide the grille out first downwards and then towards the
hands entrance, and then remove the filter.
Clean by blowing with compressed air. Replace the filter when it is dirty or damaged, using a filter
of the same type (Code RIC00041).
Failure to replace this filter can compromise operations of the equipment.
WARNING!
Lack of maintenance of the filters can compromise operation of the equipment and damage it.
Lack of maintenance of the filters can compromise equipment performance.