-257,"File name error"
. File name is empty.
-410,"Query INTERRUPTED”.
Indicates that a condition causing an INTERRUPTED
Query error occurred (see IEEE-488.2, 6.3.2.3).
1401,"Wrong program data checksum found"
. Checksum of file transferred is
invalid.
1403,"File length error"
. In file transfer length is invalid. This can be happen if there
is not enough memory or storage space on device to retrieve file.
1404,"File type error"
. Invalid file type given.
6.3
Command Reference
6.3.1
Common Commands
6.3.1.1
*CLS
Clear Status Command
The
*CLS
common command clears the status data structures by clearing all event
registers and the error queue. Also possible executing of scenario or signal generator is
stopped. It does not clear enable registers and transition filters. It clears any pending *WAI,
*OPC, and *OPC?.
Example
*CLS
6.3.1.2
*ESE
Standard Event Status Enable
Sets the enable bits of the standard event enable register. This enable register contains a
mask value for the bits to be enabled in the standard event status register. A bit that is set
true in the enable register enables the corresponding bit in the status register. An enabled
bit will set the ESB (Event Status Bit) in the Status Byte Register if the enabled event
occurs.
6.3 Command Reference
User Manual GSG-5/6 Series Rev. 27
227
Содержание GSG-5 Series
Страница 2: ......
Страница 4: ...Blank page II User Manual GSG 5 6 Series...
Страница 42: ...BLANK PAGE 2 5 Signal Power Level Considerations 26 User Manual GSG 5 6 Series Rev 27...
Страница 382: ...BLANK PAGE 6 7 Revision History SCPI Guide 366 User Manual GSG 5 6 Series Rev 27...
Страница 388: ...BLANK PAGE vi User Manual GSG 5 6 Series...